Chapter 5 — Techniques
65
The experiment is usually first done on a blank solution containing only
electrolyte, followed by a sample solution containing the substrate. The
data is collected on separate pages in Scope and the ‘blank’ data
subtracted with the Set Background command. You can then copy and
paste the scan to a spreadsheet so that the differences can be plotted
against 1/Ðt in a Cottrell graph.
For further details refer to the
Scope Software Manual
.
Chronocoulometry
Chart and Scope software have the ability to integrate an incoming
signal both online (that is, in real time as you are collecting data), or
offline (that is after the experiment is completed). Since the total charge
transferred (the total number of coulombs, or electrons, transferred) is
equal to the integrated current, the settings used for
chronoamperometry,
, can also be used for chronocoulometry.
Adjust the pulse
waveform
duration,
amplitude and
frequency
Use the Stimulator… command to access the
Stimulator (applied potential waveform) controls
Use the Output
Voltage… command
to set the baseline
potential
Figure 5–4
Scope controls for
multiple step
chronoamperometry.