Chapter 3 — The Dual Picostat
37
baseline current is being monitored click the
button. It may take
a few seconds to zero the signal. Normally this will suffice to accurately
zero the signal, and you can then choose a more sensitive current
range setting to observe your signals. Use the buttons to manually
adjust the amount of offset if required (use Ctrl-click for fine adjustment).
Each Dual Picostat channel can have a different amount of offset
applied to its current signal.
The amount of current offset is reported in the text box,
in
which you can also enter a value directly.
If you need to zero the current signals after recording has started, then
you can use the Zero All Inputs command in the Chart software Setup
menu,
.
Maintenance
The Dual Picostat will not require maintenance during daily operation.
However, you should periodically check it for optimum results.
First set up the Dual Picostat and
e-corder
as outlined earlier in this
Chapter. Connect the electrode cables connected, and attached to the
10 Mohm test resistor, as described in the section on
.
Open the Picostat Control window,
. Adjust
the gain range to 200 nA to accommodate your signal amplitude, and
select the 10 Hz low-pass filter (and/or Mains Filter) to minimise high
frequency noise on the signal — especially if you are working outside a
Faraday cage.
Figure 3–9
The Chart software Setup
menu.
Summary of Contents for e-corder
Page 62: ...58 eDAQ Potentiostats ...
Page 88: ...84 eDAQ Potentiostats ...
Page 94: ...90 eDAQ Potentiostats ...
Page 102: ...98 eDAQ Potentiostats ...