SE-1515 PC ECG User Manual ECG Analysis
- 56 -
2.
Click
Pharma
to open the
pharma stud
y
screen start pharma study.
Test Time Setting
In the parameter setup window for pharma study, you can set the report output time
mode. The default value is 0-1-3-5-10-15. It means the system will automatically output
a report in the beginning,
1st, 3rd, 5th, 10th, and 15th minute. This function is available
only when the report format for pharma study is set to
All-Lead ECG Report
.
You can also customize new time modes. A new time mode can contain a maximum of
10 time nodes and the time node value must be no larger than 30. When the cusmization
is started, you should set the time nodes first and then close the customization window.
For details about other settings, refer to corresponding parameter descriptions for resting
ECG.
3.
If you find the hand electrodes or chest electrodes have been placed incorrectly after ECG
sampling, you can click
Inversion
to adjust the electrode settings and therefore avoid
resampling.
4.
Click
Re-analyze
and the system automatically reanalyze the ECG data in the last 10s.
5.
The information displayed on the right panel includes measurement information, feature
description and diagnosis result.
Measurement information: Parameter values can be input manually. If the value is
beyond range, it will turn red automatically.
Shortcut Keyboard: Used for quick modification on feature description or diagnosis
result.
Smart input: In the
Feature Description
or
Diagnosis Result
area, you can input only
one letter and possible phrases will be provided.
Glossary: Click to open the
Glossary
window and you can edit feature description or
diagnosis result.
History: Click on
History
and you can view all the history diagnosis records of the
current patient.
6.
Right-click on the waveform area and a shortcut menu will be displayed. Options provided
are
R-R (bpm)
,
R-R (ms)
and
Refilter
.
Select
R-R (bpm)
/
R-R (ms)
and related data will be displayed.
Select
Refilter
and you can modify the parameter values in the
Refilter
window.
7.
Double-click on the waveform area, you can view the magnified waveform around the click
point on the magnified waveform screen.
Click on the lead symbol in the 1mV calibration mark and you can switch leads.
The inverted triangles correspond to R waves. Click on an inverted triangle, 5 mark lines
will appear around the corresponding R wave and related R, QRS, PR, and QT/QTC
information will be displayed. Right click and you can disable the mark lines.