SE-1515 PC ECG User Manual ECG Sampling
- 53 -
4.4.3.7 Protocol Editing
In the
Device
window, select a protocol and click
Edit
to enter the editing window for
non-default protocols. In this window, you can change any parameter of each stage or delete the
non-default protocols.
4.4.3.8 Parameter Setup
Item
Description
Target HR
You can configure the Max predicted HR and the calculation method of
the target HR.
Normal BP range
You can configure the systolic and diastolic to the value in the normal
range.
If the blood pressure of the patient excesses the configured value, an
indication message will be displayed on the main screen and the
background color of the blood pressure parameter will be yellow.
POST J
It can be set to
Manual
or
Auto
.
NOTE
: J point is the endpoint of the QRS waveform group and
the startpoint of the ST segment. It is also the reference
point used by the system to determine the place of the
ST segment. Please select the J point based on the
actual ECG waveforms of the patient.
ST Standards
When
ST Overrun hint
is selected, the system will display an
indication message
ST Overrun
when the ST value is detected to be
beyond the threshold.
You can set the threshold for
ST Elevation
or
ST Depression
only
after
STOverrun hint
is selected.
ST Elevation
can be set to
0.05 ~0.3mV,
and
ST Depression
can be set
to -
0.05 ~-0.3mV
.
Manual Report
Options provided are
Print, Print and Save, Save.
When
Print/Print and Save/Save
is selected, you can click
to
Print/Print and Save/Save
the corresponding ECG report during the
exercise test.
Auto Report
Options provided are
OFF, Print, Print and Save, Save.
If
OFF
is selected, the 12-lead report will not be printed automatically
on the scheduled auto-print time (configured on the
Edit Protocol
window) during the exercise test.