SE-1010 PC ECG User Manual EMC Information
- 78 -
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM –
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and SE-1010 PC ECG
SE-1010 PC ECG is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of SE-1010 PC ECG can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and SE-1010 PC ECG as recommended below,
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
(m)
Rated
maximum
output power of
transmitter
(W)
150 kHz to 80 MHz
P
d
2
.
1
=
80 MHz to 800 MHz
P
d
2
.
1
=
800 MHz to 2.5 GHz
P
d
3
.
2
=
0.01
0.12 0.12 0.23
0.1
0.38 0.38 0.73
1
1.2 1.2 2.3
10
3.8 3.8 7.3
100
12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.