SE-1010 PC ECG User Manual Hint Information
- 63 -
Chapter 9 Hint Information
Hint information and the corresponding causes provided by the system are listed as follows.
Table 9-1 Hint Information and Causes
Hint Information
Causes
Lead off: X
Electrodes fall off the patient or the patient
cable falls off the ECG sampling box.
It is pre-sampling now, please click on
‘Start’ to begin recording.
During the pre-sampling course
Resting ECG is sampling now!
During the sampling course of Resting ECG
Hint: Please make sure the USB line has
been connected. If possible, please
re-connect it!
The USB cable is disconnected or the
communication between the ECG sampling box
and the serial port is interrupted.
1.
Reconnect the ECG sampling box to the
PC.
2.
Click on the
Device
tab in the
System
Setting
window of the PC ECG system,
and check whether the sampling device is
set correctly.
Communication error! Please check the
USB cable!
The USB cable falls off the PC during the
sampling process.
It is connecting now, please wait…
DX12 transmitter is connecting to DX12
receiver.
Can't find the corresponding Bluetooth
sampling device, please make sure the
device is on.
Fail to connect with DX12 receiver.
Battery of sampling device is weak,
please change the battery after the test.
Battery of DX12 transmitter is low.
Battery is weak, the sampling device is
closing.
Battery of DX12 transmitter is low.
Sampling Device is in sleep mode, please
press “Power” to wake it up.
DX12 transmitter is in sleep mode.
Overload
The direct current offset voltage on an electrode
is too high.