SE-1010 PC ECG User Manual Configuring the System
- 50 -
Item Description
Filter Setting
♦
DFT Filter:
DFT filter greatly reduces the baseline fluctuations
without affecting ECG signals.
Choose from:
Weak
and
Strong
.
NOTE
: If DFT filter is set to
Strong
, the ECG data displayed on
the screen is 0.85 seconds later than the real-time ECG
data; if DFT filter is set to
Weak
, the ECG data
displayed on the screen is 1.8 seconds later than the
real-time ECG data.
♦
EMG Filter:
EMG filter suppresses the disturbance caused by strong
muscle tremor.
Choose from:
25Hz
,
35Hz
or
45Hz
.
♦
Lowpass Filter:
Lowpass filter
restricts the bandwidth of input
signals. All the input signals whose frequency is higher than the
setting cutoff frequency will be attenuated.
Choose from:
75Hz
,
100Hz
or
150Hz
♦
AC Filter:
AC filter suppresses AC interference without attenuating
or distorting ECG signals.
Choose from:
50Hz
and
60Hz
.
Sampling Time
Set the sampling time of
Resting ECG
to 10-600s.
Others
♦
Lead Sequence
Set
Lead sequence
to
Standard
or
Cabrera
, and the lead groups are
displayed or printed in the corresponding sequence listed in the
following table.
Lead
Sequence
Lead
group 1
Lead
group 2
Lead
group 3
Lead
group 4
Standard
І
, II,
ІІІ
aVR, aVL,
aVF
V1, V2,
V3
V4, V5,
V6
Cabrera
aVL,
І
,
-aVR
II, aVF,
ІІІ
V1, V2,
V3
V4, V5,
V6
♦
HR Calculation
Set the
HR calculation
to one of the 12 standard leads:
І
,
ІІ
,
ІІІ
,
aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6.
♦
HR Analysis Lead
Set the
HR Analysis Lead
to one of the 12 standard leads:
І
,
ІІ
,
ІІІ
,
aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6.