EPX-SERISE USER MANUAL Rev. G
28
4.7.2 SPECIFICATION TABLE
4.7.2.1 EPX-F1600
Max Output Power
1.6 kW
Input Power
Voltage
200V-240V
Phase & Frequency
Single / 50/60 Hz
Radiography kV Range in 1kV steps
(40KV ~ 90KV)
Tube voltage
current
mAs
40KV ~ 50KV
30mA
0.4 ~ 20
40KV ~ 50KV
25mA
25~32
51KV ~ 60KV
25mA
0.4 ~ 20
51KV ~ 60KV
20mA
25~32
61KV ~ 70KV
20mA
0.4 ~ 20
61KV ~ 70KV
15mA
25~32
71KV ~ 80KV
18 mA
0.4 ~ 20
71KV ~ 80KV
14 mA
25~32
81KV ~ 90KV
16 mA
0.4 ~ 20
81KV ~ 90KV
12 mA
25~32
MAs Range
0.4mAs
– 32mAs, 20steps
Max. kV Deviation
±3 %
Max. mAs Deviation
±5 %
Max. Accuracy of Loading Time
±13%
Max. Reproducibility of the RADIATION output
Coefficient of variation < 0.01
Display
KV /mAs: 7-segment LED
X-ray Tube
Model Name
D-125 (Toshiba)
Focal Spot
1.2mm x 1.2mm
Target Angle
16 degree
Anode Heat Storage
50 KHU
Total Filtration
Tube:0.8mmAl / collimator: 0.5mmAl / filter:2.0mmAl
3.3 mm Al eq. @ 75kV
Collimator
with Laser
Pointer
Type
Double slit type, manual operation
Min. X-ray Field Size
0cm x 0cm @100cm SID
Max X-ray Field Size
36cm x 36cm @ 100cm SID
Laser Pointer
Class II
Timer
Push button illuminator with 30 sec timer
Lamp
24V, 50W Halogen
Weight
13 Kg