EPX-SERISE USER MANUAL Rev. G
51
6.1.1 Normal Mode
a. Connect power plug into the device.
b. Turn on the power switch located on the back of the equipment.
c. Select the KV values using the KV up/down switch.
d. Select the mAs values using the mAs up/down switch.
e. Positioning the machine at the ready condition by pushing the hand one time switch.
f. When the
“Ready LED” lighted, press the hand switch again and make exposure. During the x-ray
exposure,
“X-ray LED” will be lighted on.
g.
After the exposure is ended, “wait LED” sign will be lighted on for the waiting time and beep sound
will be heard.
6.1.2 APR Mode
a. Connect power plug into the device.
b. Turn on the power switch.
c. Select the APR memory out of 1~6.
d. Select the kV values using the KV up/down switch.
e. Select the mAs values using the mAs up/down switch.
f. Save the selected K
V and mAs figures with pushing the “save” button.
g. When the
“Ready LED” sign appears, press the hand switch again and make exposure. During the
x-ray exposure, the
“X-ray LED” will be lighted on.
h
. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time and beep sound
will be heard.
6.1.3 Dosimetry Data Display Mode
a. After the exposure is ended, dosimetry data will be displayed on the FND for two seconds.
b. Push the button 1 for two seconds, there will be displayed a dosimetry data for current kV and mAs
value.
c. Push the button 2 for two seconds, there will be displayed the total exposed dose.
d. Push the button 3 for two seconds, there will be reset to zero.