A short press of the MODE button after audio signaling of every registered
gamma quantum has been switched on or off, fixes all programmed values in the
nonvolatile memory of the radiometer. It also finishes the submode of new values
programming of alarm threshold level and specified statistical error, as well as
switching on/off audio signaling of registered gamma-quanta.
If the programmed values are fixed, a new value of the threshold level blinks
three times on the LCD, and the radiometer returns to the mode of photon-ionizing
radiation DER measurement.
Caution!
If the submode of new values programming of alarm threshold level
and specified statistical error, as well as switching on/off audio signaling of
registered gamma-quanta is paused for more than 30 s (the user presses no buttons
of the radiometer), the radiometer will automatically return to the mode of photon-
ionizing radiation DER measurement. All changes made in the submode of new
values programming will be cancelled.
Summary of Contents for STORA-TU RKS-01
Page 1: ...RKS 01 STORA TU GAMMA BETA RADIOMETER DOSIMETER Operating manual...
Page 2: ......
Page 20: ...18 Figure 1 External view of the radiometer top view...
Page 22: ...20 Figure 2 External view of the radiometer bottom view...
Page 38: ...36 Figure 3 LCD of the radiometer the mode of photon ionizing radiation DER measurement...
Page 42: ...40 Figure 4 LCD of the radiometer viewing the value of specified statistical error...
Page 46: ...44 Figure 7 LCD of the radiometer the submode of alarm threshold level programming...
Page 59: ...57 Figure 12 LCD of the radiometer the mode of surface beta particles flux density measurement...
Page 63: ...61 Figure 13 LCD of the radiometer viewing the value of specified statistical error...
Page 67: ...65 Figure 16 LCD of the radiometer the submode of alarm threshold level programming...
Page 73: ...71 Figure 19 LCD of the radiometer the clock mode...
Page 76: ...74 Figure 21 LCD of the radiometer the submode of time and date correction year programming...
Page 78: ...76 Figure 22 LCD of the radiometer the submode of time and date correction date programming...
Page 81: ...79 Figure 23 LCD of the radiometer the alarm clock mode...
Page 82: ...80 Figure 24 LCD of the radiometer the submode of alarm clock activation time programming...
Page 87: ...85 Figure 26 LCD of the radiometer the mode of data communications control with the PC...
Page 89: ...87 Figure 27 LCD of the radiometer the mode of data communications control with the PC...
Page 94: ...92 Figure 29 LCD of the radiometer measurement results viewing...