11
1.2.3 The radiometer measures photon-ionizing radiation DER and surface
beta-particles flux density until the specified statistical error is gained.
1.2.3.1 The specified statistical error can be programmed by the user or
automatically determined by the radiometer depending on the radiation intensity.
1.2.4 For fast of photon-ionizing radiation DER and surface beta-particles flux
density the radiometer provides a twelve-segment indicator of instantaneous value.
The information is updated on the indicator of instantaneous value each 500 ms.
1.2.5 Threshold alarm system with two independent threshold levels is realized
in the radiometer:
- photon-ionizing radiation DER;
- surface beta-particles flux density.
1.2.5.1 Threshold level values of photon-ionizing radiation DER are
programmed in the range of 0 to 999,9 µSv/h with discreteness of 0.01 µSv/h.
1.2.5.2 Threshold level values of surface beta-particles flux density are
programmed in the range of 0 to 9999·10
3
part./(cm
2
·min) with discreteness of
0.01
10
3
part./(cm
2
·min).
Summary of Contents for STORA-TU RKS-01
Page 1: ...RKS 01 STORA TU GAMMA BETA RADIOMETER DOSIMETER Operating manual...
Page 2: ......
Page 20: ...18 Figure 1 External view of the radiometer top view...
Page 22: ...20 Figure 2 External view of the radiometer bottom view...
Page 38: ...36 Figure 3 LCD of the radiometer the mode of photon ionizing radiation DER measurement...
Page 42: ...40 Figure 4 LCD of the radiometer viewing the value of specified statistical error...
Page 46: ...44 Figure 7 LCD of the radiometer the submode of alarm threshold level programming...
Page 59: ...57 Figure 12 LCD of the radiometer the mode of surface beta particles flux density measurement...
Page 63: ...61 Figure 13 LCD of the radiometer viewing the value of specified statistical error...
Page 67: ...65 Figure 16 LCD of the radiometer the submode of alarm threshold level programming...
Page 73: ...71 Figure 19 LCD of the radiometer the clock mode...
Page 76: ...74 Figure 21 LCD of the radiometer the submode of time and date correction year programming...
Page 78: ...76 Figure 22 LCD of the radiometer the submode of time and date correction date programming...
Page 81: ...79 Figure 23 LCD of the radiometer the alarm clock mode...
Page 82: ...80 Figure 24 LCD of the radiometer the submode of alarm clock activation time programming...
Page 87: ...85 Figure 26 LCD of the radiometer the mode of data communications control with the PC...
Page 89: ...87 Figure 27 LCD of the radiometer the mode of data communications control with the PC...
Page 94: ...92 Figure 29 LCD of the radiometer measurement results viewing...