106
Calculate the DER value in µSv/h by the formula (3.1).
Calculate the main relative error of measurement in percentage.
Place the УПГД-3В carriage together with the radiometer in the position,
where DER from
137
Cs source is 800 µSv/h, and wait until the statistical error of
DER measurement results goes down to the value of not more than 10 %. Then
register five DER measurement results in the protocol with 5 s interval.
Calculate the average DER value and the main relative error of measurement
in percentage.
The radiometer is acknowledged to have passed the verification if the main
relative error in percentage during measurement of each DER level does not exceed
)
10
(
2
15
H
, where
)
10
(
H
is a numeric value of the measured DER in µSv/h.
Summary of Contents for STORA-TU RKS-01
Page 1: ...RKS 01 STORA TU GAMMA BETA RADIOMETER DOSIMETER Operating manual...
Page 2: ......
Page 20: ...18 Figure 1 External view of the radiometer top view...
Page 22: ...20 Figure 2 External view of the radiometer bottom view...
Page 38: ...36 Figure 3 LCD of the radiometer the mode of photon ionizing radiation DER measurement...
Page 42: ...40 Figure 4 LCD of the radiometer viewing the value of specified statistical error...
Page 46: ...44 Figure 7 LCD of the radiometer the submode of alarm threshold level programming...
Page 59: ...57 Figure 12 LCD of the radiometer the mode of surface beta particles flux density measurement...
Page 63: ...61 Figure 13 LCD of the radiometer viewing the value of specified statistical error...
Page 67: ...65 Figure 16 LCD of the radiometer the submode of alarm threshold level programming...
Page 73: ...71 Figure 19 LCD of the radiometer the clock mode...
Page 76: ...74 Figure 21 LCD of the radiometer the submode of time and date correction year programming...
Page 78: ...76 Figure 22 LCD of the radiometer the submode of time and date correction date programming...
Page 81: ...79 Figure 23 LCD of the radiometer the alarm clock mode...
Page 82: ...80 Figure 24 LCD of the radiometer the submode of alarm clock activation time programming...
Page 87: ...85 Figure 26 LCD of the radiometer the mode of data communications control with the PC...
Page 89: ...87 Figure 27 LCD of the radiometer the mode of data communications control with the PC...
Page 94: ...92 Figure 29 LCD of the radiometer measurement results viewing...