Lead Fault Signal Display Setting
When there is a Lead Fault event, this setup sets whether the “Lead Fault” message
will be displayed on the LCD screen of the device or not.
Selecting [L/F] menu will switch between ON and OFF in the menu, with ON turning
on the message display and OFF switching off the message display.
The following message for Lead Fault will be shown with affected channels:
Note
-
If the RL lead becomes faulty, the device will display a message saying all
leads are faulty (“Lead Fault! All Leads.”)
-
If a Lead Fault occurs during monitoring or recording, a buzzer alarm will
sound and the message box will pop up.
-
No pacemaker signal may be detected. Additionally, if a Lead Fault occurs,
please re-measure the EKG as it could have affected the measurement by
the device.
Lead Fault! All Leads.