The following adjustments are available in the Adjustments / Internal / Values options menu:
Parameter
Meaning
Adjustment
Factory setting
Temperature
Not available at all instruments
°C or °F
°C
Excess air
Excess Air as proportionate
number or in %
proportionate number or %
proportionate number
Eff.(C)
Efficiency calculation with or
without condensation gain
On, Off
On
Draught
Unit of draught measurement
hPa, mmH2O, “H2O, PSI, mbar hPa
The following adjustments are available in the Adjustments / Internal / USB menu:
Parameter
Meaning
Adjustment
Factory setting
Transfer speed
1200 … 115200 Baud
38400 Baud
Protocol
DAS = DAS NT 2
Enhanced = DAS 5
---- = no protocol
Enhanced
The following settings are predefined in the Settings / Internal / Bluetooth (BLE) menu and can
NOT be changed:
Parameter
Meaning
Adjustment
Factory setting
Transfer speed
115200 Baud
115200 Baud
Protocol
Enhanced
Enhanced
The following adjustments are available in the Adjustments / Internal / WLAN menu:
Parameter
Meaning
Adjustment
Factory setting
Access Point
OR
Existing Network
Set whether the instrument is to
be used as a WiFi access point
OR
to connect to an existing WiFi
network
Access Point,
Existing Network
Access Point
Hint
The Access Point setting is necessary to connect the ecom-EN3 to mobile
devices (smartphones, tablets) via the Wi-Fi interface.
Activate "Access Point" to connect the ecom-EN3 with mobile
devices via WLAN / Wi-Fi.