6.4. Pressure check
The pressure check (long term measurement up to
1500 hPa or 20 bar with external pressure sensor) is
registered as a control pragram in the
ecom-DP3
.
Proceed as follows:
1. Scroll with the cursor keys
<up/down>
in the
menu
"More measurements"
to the menu point
„Pressure check“
. Activate the menu point with
<Enter>
.
2. Check if necessary the values of stabilisation time,
measurement time and test pressure with
<F1>
(back with
<ESC>
). If the measurement time is set to
0, thecontrol program runs until it is interrupted.
3. Lock the gas pipe with a suitable adapter (test
stopple, high pressure stopple or one-pipe counter
cap) or connect the external pressure sensor.
4. Connect the components as described in
chapter 3.
5. Start the control program with
<Enter>
and built
up the adjusted pressure.
6. If the internal sensor (up to 1500 hPa) is used, the
needed pressure can be built up with the internal
pump (switch on and off with
<F3>
).
7. Wait for stabilisation time. If necessary the stabili-
sation time can be terminated with
<F1>
.
8. Once the measurement time is over (or with
<En-
ter>
if measurement time = 0), the result is displayed
and can be printed with
<Print>
.
9. If the menu
„Pressure check“
is selected again, so the result can be called
up with
<F4>
(as long as the
ecom-DP3
is on) or a new measurement can be
started with
<F1>
.
Pressure check
1,5Bar/1500hPa long tes
Stabilisation time
Res. time: 58sek
Pressu 1003.1hPa -0.8hPa
Temp. 22.6°C 0.3°C
Start Difference
1002.3
hPa
Pressure check
1,5Bar/1500hPa long tes
Measur. in process
Res. time: 9:25min
Pressu 1002.3hPa 0.0hPa
Temp. 22.6°C 0.0°C
Start Difference
1002.3
hPa
Pressure check
1,5Bar/1500hPa long tes
Measurement 10:00min
Ini. Press 1002.3hPa
Fin. Press 1002.3hPa
Difference 0.0hPa
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
0.0°C
Forward w.:
Pressure check
1,5Bar/1500hPa long tes
Start new
measurement?
Select:
YES
NO
Pressure check
1,5Bar/1500hPa long tes
Build up pressure
Forward w.:
0.0
hPa
Summary of Contents for DP3