4
INDEX
SECTION DESCRIPTION PAGE
Notice and Warranty Policy 3
1.0 General 5
2.0 Specifications 5-6
3.0 Installation 6
3.1 Installation “Feet” Drawings 6
4.0 System Enclosure Exterior Dimensional Drawing 7
4.1 System Enclosure Interior Layout Drawing 8
4.2 Wiring 8-9
4.3 Wiring Terminals Strip Photos 9
4.4 Wiring Connections Drawing 10
5.0 Main Circuit Board Photo 11
6.0 System Operation 12-13
7.0 System Programming - General 14
7.1 System Programming - Input Codes 14
7.2 System Programming - Output Codes 15
7.3 System Programming - Analog Output Codes 15
7.4 Other Codes 15
7.5 Programming Push-Buttons Photo 16
7.6 System Programming - Input Code Description 16-20
7.7 System Programming - Relay Output Code Description 21-23
7.8 System Programming - Analog Output Code Description 24-25
7.9 System Programming - Other Codes 26
8.0 Replacement Parts 26
25
7.8 SYSTEM PROGRAMMING - ANALOG OUTPUT CODES, CONT’D…..
“1331” Calibrate Analog Output Low Limit:
This code provides the user with a
method to calibrate the low limit for each analog output channel. An accurate digital
mulit-meter is required. First attach meter leads to channel-1 output terminal strip.
Set the meter to the DC mA scale. This is a locked function. To proceed, first input
the software unlock code. Next, input code “1331”. The voltage output will be
displayed on your digital multi-meter.
To calibrate this channel, depress push-button “3” to increment to “1” to decrement
the value to achieve “4.00” mA on your meter. Depress push-button “2” to scroll to
the next channel. Move only the positive meter lead to channel-2 and repeat the last
step to once again achieve a reading of 4.00 mA on your meter. Repeat the last two
steps for the last two channels, if they have been enabled for use. Once this function
is finished, continue to depress push-button “2” until you reach the last channel and
the system will automatically start to scroll through all “active” channels.
Selection: N/A
“1332” Calibrate Analog Output High Limit:
This code provides the user with a
method to calibrate the high limit for each analog output channel. An accurate
digital mulit-meter is required. First attach meter leads to channel-1 output terminal
strip. Set the meter to the DC mA scale. This is a locked function. To proceed, first
input the software unlock code. Next, input code “1332”. The voltage output will be
displayed on your digital multi-meter.
To calibrate this channel, depress push-button “3” to increment to “1” to decrement
the value to achieve “20.00” mA on your meter. Depress push-button “2” to scroll to
the next channel. Move only the positive meter lead to channel-2 and repeat the last
step to once again achieve a reading of 20.00 mA on your meter. Repeat the last
two steps for the last two channels, if they have been enabled for use. Once this
function is finished, continue to depress push-button “2” until you reach the last
channel and the system will automatically start to scroll through all “active” channels.
Selection: N/A
AC 1 Menu 1331
Null
AC 1 Menu 1332
Span
Summary of Contents for PAC-44 Series
Page 2: ...2 27...