background image

 INMANUUIN182

5

 

DESCRIPTION 

FRONT PANEL 

 

1

 

On/Off switch 

2

 

Diaphragm control 

3

 

Camera connector (video input) 

4

 

Universal fiberoptic cable connector 

 

 

 

Summary of Contents for XENOLUX 180

Page 1: ...XENOLUX 180 Light Source V 2019 04 01...

Page 2: ...ECLERIS USA 11801 NW 100th Rd Suite 17 Medley FL 33178 Phone 1 786 483 7490 CMC Medical Devices Drugs S L C Horacio Lengo N 18 CP29006 M laga Spain Ph 34951214054 Email info cmcmedicaldevices com...

Page 3: ...IFICATIONS 7 SIMBOLS DESCRIPTION 8 COMPONENTS 9 INSTALLATION 10 REVISION AND OPERATION 11 REVISION 11 MAINTENANCE 13 CLEANING 13 External Surfaces 13 ELECTRICAL CONNECTIONS 14 FUSE CHANGE 14 LAMP CHAN...

Page 4: ...M universal endocamera which has a 1 3 CCD sensor and integrated controls especially designed for medical applications ensuring a high image quality precision and resolution This endocamera allows the...

Page 5: ...INMANUUIN182 5 DESCRIPTION FRONT PANEL 1 On Off switch 2 Diaphragm control 3 Camera connector video input 4 Universal fiberoptic cable connector 1 3 4 2...

Page 6: ...tor with fuse holder 7 S Video Output SVHS 8 Composite Video Output CVBS 9 EclerisEndodigiConnector Image and video capture and other peripheral devices 10 Ventilationgrids Standard output Normally Op...

Page 7: ...MPERATURE 6000 K Approx REFRIGERATION Forced through fans FIBER OPTIC CONNECTOR Built in revolving turret with multiple adapters VIDEO OUTPUT Composite Video CVBS 1 Vpp 75 BNC S Video Y C Jack mini DI...

Page 8: ...SIMBOLS DESCRIPTION Attention Read the related documentation Alternate current Protect against water Up Fragile CE Mark Identificaci n del fabricante Nombre Direcci n B Type...

Page 9: ...2 9 COMPONENTS XENOLUX 180 light source includes the following components It is recommended to identify and check their existence before proceeding with the installation MainUnit PowerSupply Cable Use...

Page 10: ...the Ecleris Microscope Colposcope forearm See Light Source Installation Section in Ecleris Microscope Colposcope User s Manual 6 Rotate the universal adapter until the Wolf connector is aligned with t...

Page 11: ...be made before turning the equipment on The XENOLUX 180 shall only be used in places with electrical installations that meet the national provincial and local regulations in effect If before connectin...

Page 12: ...Use the diaphragm control to regulate lighting intensity 4 Once the procedure is over turn the XENOLUX 180 off in order to maximize the lamp useful life 5 Should you use a video camera Ecleris PROCAM...

Page 13: ...n charge CLEANING Before cleaning the XENOLUX 180 disconnect the power supply cable form the electrical power External Surfaces Clean the external surface of the equipment using a piece of cotton or c...

Page 14: ...tic holder where the fuses are located Using a screwdriver and with a slight pressure remove the lid to replace the damaged fuse 3 Remove the burnt fuse and replace it with the spare fuse located insi...

Page 15: ...ace the new 180W Xenon lamp in the lamp holder as shown in the figures code part XBO R180 45C 6 Fix the lamp by adjusting the lamp holder screw 7 Connect the Xenon lamp power supply connector male to...

Page 16: ...place or deficiently connected Remove the lamp from the lamp holder and place it again correctly High temperature Ventilationgrids are obstructed Cleantheventilationgrids A fan is not working Turn the...

Page 17: ...utral soap or any specific product to clean electronic equipment cabinets Do not introduce objects in the holes or rest any containers with liquid on the equipment Prevent the power supply cable from...

Page 18: ...r manipulation by unauthorized personal any material or personal damage were caused to third parties ECLERIS the manufacturer and the supplier of the equipment shall not assume any responsibility for...

Page 19: ...his card shall only be completed by the authorized person from the company responsible for the maintenance repair and modification of the device We recommend keeping it updated for a quicker and more...

Page 20: ...r or any ECLERIS office 2 Give detailed information of the problem 3 f the problem cannot be solved by troubleshooting the product shall be sent for its repair The following product information shall...

Page 21: ...caused to by the XENOLUX 180 due to incorrect use negligence cleaning or incorrect storage Any repairs made by personnel not authorized by ECLERIS will void this warranty The described warranty is exc...

Page 22: ...roduced by any means photocopy microfilm or other methods without the previous written authorization by ECLERIS likewise no part shall be reproduced multiplied or published through electronic data pro...

Reviews: