TROUBLESHOOTING
The following table indicates the most common problems and their quick solutions.
Please consult this table for any difficulty you may encounter.
PROBLEM
PROBABLE CAUSE
ACTION
The equipment
does not turn on
It is not connected to the
electrical power
Connect the power supply cable
to the electrical power
The power supply cable is
broken or cut
Change the power supply cable
The plug is in bad condition
Changetheplug
Fuse brokenorburnt
Changethe fuse
The green pilot
light from the
on/off switch turns
on but the lamp
does not
Exhaustelamp
Changethelamp
Internal fuse burnt or
defective board
Contact the closest Technical
service
Light intensity is
not the correct one
Defectivelamp
Changethelamp
The lamp is out of place or
deficiently connected.
Remove the lamp from the
lamp-holder and place it again
correctly.
High temperature
Ventilationgrids are
obstructed
Cleantheventilationgrids
A fan is not working
Turn the equipment off
immediately and contact the
closest Technical Service
Excessivenoise
Lamp cables are touching
the fan
Arrange the lamp cables in a
way they do not touch the fan
Light is correct but
no image appears
on the monitor
when connecting
the video camera
The video cable is cutor not
connected to the monitor
Change the cable and check the
connection to the monitor
Defective camera
Replacethe video camera
Deficientlyprogrammed
monitor
Select the corresponding video
input and Standard
Imagetoodark
Inadequatelighting
Open the diaphragm and check
that the lamp and fiber optic
cable are correctly placed
Deficientlyprogrammed
monitor
Check the monitor color,
brightness and saturation levels
after having made a white
balance.