REVISION AND OPERATION
REVISION
Once installation is over and before turning the XENOLUX 180 on, please go through
the following check list:
•
Verify that the electrical power has its grounding connection in perfect
conditions and that it makes perfect contact with the equipment.
•
Check that the fuse holder has a fuse installed and that it is the correct one
(T5AL250V for 110V power supply and T2.5AL250V for 220 V).
•
Check that the ventilation grid is completely free, without obstruction.
•
Ensure an adequate air flow.
•
Verify that there are no water or liquid sources near the equipment that may
damage it, such as gases or flammable emissions in the environment.
All connections shall be made before turning the equipment on.
The XENOLUX 180 shall only be used in places with electrical
installations that meet the national, provincial and local regulations in
effect.
If before connecting the XENOLUX 180 any anomaly is detected, please contact the
technical service department of your local distributor or the closest ECLERIS office,
prior to its use.
OPERATION
This section thoroughly describes the equipment start-up and functioning verification
procedure. It is recommended to perform this procedure after installing the equipment
and during maintenance inspections.
The equipment shall only be operated by qualified personnel.
Once the XENOLUX 180 is turned on, it shall remain in this condition
for at least 15 minutes.
When turning the XENOLUX 180 off, wait for at least other 15 minutes
before turning it on again.
Short On and Off-cycles considerable reduce the lamp useful life.