2. SAFETY INFORMATION
This manual contains all the relevant information on the appropriate care and handling of the device
while on operation, as well as information about its accessories.
This symbol shall be used in the manual to indicate information that includes important
warnings or precautions.
Read this manual carefully before using the device.
Should you have any doubt regarding the contents of the manual herein, or the operation and safety
of the device, please contact your local distributor or the nearest ECLERIS office.
ECLERIS will not be held responsible for damage that can occur to the equipment due to not
following the instructions or not observing the following criteria:
Keep the equipment out of children reach
Do not install the equipment next to heat sources, such as heaters.
Install the equipment on a firm horizontal surface, in a well-ventilated area.
Protect the equipment and accessories from direct solar light, dust, humidity and vibrations
or excessive banging.
Equipment not appropriate to be used near or around flammable anesthetic mixture with air,
oxygen or nitrous oxide.
The equipment does not need sterilization.
Do not put any container with liquids on top of the device.
Main power cord cable are resistant to deformations and mechanical tensions however do
not damage or avoid violent shocks to the cables.
Do not use the equipment during electrical storms.
Unplug the equipment when it is not being used.
This equipment has to be connected to ground through a ground conductor in the power
cable. This ground connection is essential for the safe use of the equipment.
Do not connect the equipment to devices that do not comply with IEC60601-1ed 3.0 or later.
The device is provided with a disconnectable power cable that assures proper electrical
isolation. If it needs to be replaced, the replacement should have the same features:
3x0.75mm
2
.
In case of interference with other electro medical or electric equipment, we recommend
increasing the distance between PROCAM HD and the other equipment to minimize these
effects.
Do not open the equipment. Maintenance and repairs must be done by ECLERIS or an
authorized company. ECLERIS will not be responsible for the consequences of repairs or
maintenance carried out by non-authorized personal.