![Ecleris MINIVAC+ User Manual Download Page 11](http://html1.mh-extra.com/html/ecleris/minivac/minivac_user-manual_3952453011.webp)
USER MANUAL
–
Page
11
of
18
INMANUUIN155 Rev.2019.02.14 / ENG
The equipment emits a
series of
beep
at the
end of the cleaning
cycle.
Do not use hot water.
1. To begin the cleaning cycle, the vacuum circuit and the infusion circuit must be stopped,
and the applicator must have a diamond tip.
2. Place the applicator in a vertical position, embedded in the receptacle for the applicator.
3. Place the cleaning cap on the applicator so that the residues are properly aspirated.
4. Place a solution of 70% ethyl alcohol and 30% distilled water in the receptacle for the
products.
5.
Press and hold the button
for 5 seconds to activate the cleaning cycle
.
To perform the cleaning cycle the cap must be placed on the diamond tip. Do not perform the
cleaning cycle without attaching the cleaning cap. Do not install the cleaning cap if the
applicator does not have a diamond tip
.
You must keep the button
pressed for 5 seconds, if you just press it, the cleaning cycle
will not start.
Cleaning cap
Alcohol 70%