7.  TECHNICAL SPECIFICATIONS 

POWER SUPPLY VOLTAGE 

100 -240V 

Nº OF PHASES 

1 (single phase) 

FREQUENCY 

50 

– 60 Hz 

PROTECTION AGAINST ELECTRICAL SHOCK 

Class I 

APPLIED PART 

BF 

PROTECTION LEVEL AGAINST WATER 

ENTRANCE 

IPX0 

CONSUMPTION 

110VA 

FUNCTION MODE 

Continuous 

FUSES 

2 x T3AL250V 

COLOR TEMPERATURE 

6000ºK approx. 

LED SYSTEM SERVICE LIFE 

Approx. 50000 hours 

REFRIGERATION 

Forced 

OVER TEMPERATURE PROTECTION 

Automatic LED system power-off 

DIMMER 

Electronic 

FIBER OPTIC CONNECTOR 

Universal fiber optic cable connector 
(Revolving Turret) 

VIDEO OUTPUT 

Composite video (CVBS) (1 Vpp/

75 Ω), 

BNC 
S-Video (Y/C), Jack mini-DIN 

ECLERIS ENDODIGI CONNECTOR 

Jack 3.5mm mono 

DIMENSIONS 

120x300x155mm (width x depth x high) 

MASS 

7.7 pounds 

SERVICE CONDITIONS 

50  

– 104ºF 

35 

– 75% relative humidity 

700 

– 1060 hPa 

STORAGE AND TRANSPORTATION  

CONDITIONS 

-22 

– 122ºF 

10 

– 95% relative humidity 

500 

– 1060 hPa 

8.  TROUBLESHOOTING 

The following table indicates the most common problems and their quick solutions. Please consult 
this table for any difficulty you may encounter. 

PROBLEM 

PROBABLE CAUSE 

ACTION 

The 

equipment 

does not turn on 

It is not connected to the electrical 
network 

Connect  the  power  supply  cable 
to the electrical network 

The  power  supply  cable  is  broken 
or cut 

Change the power supply cable 

The plug is in bad condition 

Change the plug 

Fuse broken or burnt 

Change the fuse 

Summary of Contents for EVERLUX

Page 1: ...ut previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified...

Page 2: ...3 2 INTENDED USER 5 3 3 FEATURES AND DESCRIPTION 5 3 3 1 FRONT PANEL 5 3 3 2 REAR PANEL 5 3 4 COMPONENTS 6 4 INSTALLATION AND SET UP 6 4 1 REVISION 7 5 OPERATION FUNCTIONS 7 6 MAINTENANCE 7 6 1 CLEAN...

Page 3: ...recautions Please read the manual thoroughly before operating the equipment Should you have any doubt regarding the contents of the manual herein or the equipment operation and safety please contact y...

Page 4: ...blem the system must be laid a plane surface and power cable disconnected The device must be used only by personnel informed of all this warnings Any use outside of the manufacturer specific intent us...

Page 5: ...its use for all types of procedures The EVERLUX allows the connection of the Ecleris PROCAM universal endocamera which has a 1 3 CCD sensor and integrated controls especially designed for medical appl...

Page 6: ...l match perfectly and not exceed the adapter 5 In case of using the EVERLUX as a Microscope Colposcope light source mount the EVERLUX on the Ecleris Microscope Colposcope forearm See Light Source Inst...

Page 7: ...S office prior to its use 5 OPERATION FUNCTIONS This section thoroughly describes the device functioning This system must be used only for medical paramedical or health professionals 1 Connect the pow...

Page 8: ...following the instruction provided by their manufacturer Do not turn the light source on until an hour 1 has passed from the cleaning since the unit can get damaged It is recommended to clean the exte...

Page 9: ...UT Composite video CVBS 1 Vpp 75 BNC S Video Y C Jack mini DIN ECLERIS ENDODIGI CONNECTOR Jack 3 5mm mono DIMENSIONS 120x300x155mm width x depth x high MASS 7 7 pounds SERVICE CONDITIONS 50 104 F 35 7...

Page 10: ...Lateral ventilation grids are obstructed Clear the lateral ventilation grids Fan is not working Turn the equipment off immediately and contact the closest Technical Service Light is correct but no im...

Page 11: ...blem description The shipment should be done in the original packaging to avoid damage to the equipment during the transportation ECLERIS will not be responsible for damage caused during the transport...

Reviews: