To  perform  any  revision  or  maintenance  work,  contact  your  local  distributor  or  the  closest 
ECLERIS office. 

6.1 CLEANING 

Before cleaning the EVERLUX, disconnect the power supply cable form the electrical network. 

6.1.1  EXTERNAL SURFACES 

Clean the external surface of the equipment using a piece of cotton or cloth moistened with mild 
detergent or soap. Use only adequate products for the external cleaning of electrical equipments, 
prepared in accordance with the manufacturer’s indications. 

  Perform  the  cleaning  with  special  care,  avoiding  water  or  humidity  access  to  the 

equipment.  

  Do not use flammable products, especially alcohol when the EVERLUX is hot. 

  Do not send to the autoclave any part or component of this device that are not specifically 

marked as autoclavable. 

  Do no sterilize or immerse the device or its accessories 

  The cleaning, disinfection and sterilization of accessories not provided by ECLERIS must 

be performed following the instruction provided by their manufacturer 

  Do not turn the light source on until an hour (1) has passed from the cleaning, since the 

unit can get damaged.  

  It is recommended to clean the exterior surface of the device every 15 days 

6.2 ELECTRICAL CONNECTIONS 

Constantly check the condition of all components and their aging process.  

Verify there are no: 

  Faulty, broken or cut cables.  

  Faulty connection between the cable and the plug. 

  Oxide corrosion in connection pins. 

6.3 FUSE CHANGE 

To avoid time losses due to downtime of the EVERLUX, it is recommended to have spare fuses. 

1)  Disconnect the power supply cable from the electrical network and the equipment. 
2)  Under the power supply cable connector there is a small rectangular plastic holder where 

the  fuses  are  located.  Using  a  screwdriver  and  with  a  slight  pressure  remove  the  lid  to 
replace the damaged fuse.  

3)  Remove the burnt fuse and replace it with the spare fuse. Get a new spare fuse to avoid 

loss of time and an inoperative unit.  
 

To avoid shock, injury, or damage to the EVERLUX, use ONLY fuses with the same 
specified ratings (T3AL250V) 

 

4) 

Place the fuse lid back pushing it until a “click” is heard”. 

5)  Connect the power supply cable. 

 

Summary of Contents for EVERLUX

Page 1: ...ut previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified...

Page 2: ...3 2 INTENDED USER 5 3 3 FEATURES AND DESCRIPTION 5 3 3 1 FRONT PANEL 5 3 3 2 REAR PANEL 5 3 4 COMPONENTS 6 4 INSTALLATION AND SET UP 6 4 1 REVISION 7 5 OPERATION FUNCTIONS 7 6 MAINTENANCE 7 6 1 CLEAN...

Page 3: ...recautions Please read the manual thoroughly before operating the equipment Should you have any doubt regarding the contents of the manual herein or the equipment operation and safety please contact y...

Page 4: ...blem the system must be laid a plane surface and power cable disconnected The device must be used only by personnel informed of all this warnings Any use outside of the manufacturer specific intent us...

Page 5: ...its use for all types of procedures The EVERLUX allows the connection of the Ecleris PROCAM universal endocamera which has a 1 3 CCD sensor and integrated controls especially designed for medical appl...

Page 6: ...l match perfectly and not exceed the adapter 5 In case of using the EVERLUX as a Microscope Colposcope light source mount the EVERLUX on the Ecleris Microscope Colposcope forearm See Light Source Inst...

Page 7: ...S office prior to its use 5 OPERATION FUNCTIONS This section thoroughly describes the device functioning This system must be used only for medical paramedical or health professionals 1 Connect the pow...

Page 8: ...following the instruction provided by their manufacturer Do not turn the light source on until an hour 1 has passed from the cleaning since the unit can get damaged It is recommended to clean the exte...

Page 9: ...UT Composite video CVBS 1 Vpp 75 BNC S Video Y C Jack mini DIN ECLERIS ENDODIGI CONNECTOR Jack 3 5mm mono DIMENSIONS 120x300x155mm width x depth x high MASS 7 7 pounds SERVICE CONDITIONS 50 104 F 35 7...

Page 10: ...Lateral ventilation grids are obstructed Clear the lateral ventilation grids Fan is not working Turn the equipment off immediately and contact the closest Technical Service Light is correct but no im...

Page 11: ...blem description The shipment should be done in the original packaging to avoid damage to the equipment during the transportation ECLERIS will not be responsible for damage caused during the transport...