Ecleris EVERLUX User Manual Download Page 7

USER MANUAL 

– EVERLUX 

 

Page 

7

 de 

11

 

INMANUUIN167 Rev.2019.04.01 / ENG 

4.1 REVISION 

Once  installation  is  over  and  before  turning  the  EVERLUX  on,  please  go  through  the  following 
check list: 

  Verify  that  the  electrical  network  has  its  grounding  connection  in  perfect  conditions  and 

that it makes perfect contact with the equipment. 

  Check that the fuse holder has a fuse installed and that it is the correct one (T3AL250V). 

  Check that the lateral ventilation grid is completely free, without obstruction. 

  Ensure an adequate air flow. 

  Verify  that  there  are  no water  or  liquid  sources  near  the  equipment that  may  damage  it, 

such as gases or flammable emissions in the environment. 

 All connections shall be made before turning the equipment on. 

The  EVERLUX  shall  only  be  used  in  places  with  electrical  installations  that  meet  the 
national, provincial and local regulations in effect. 

If before connecting the EVERLUX any anomaly is detected, please contact the technical service 
department of your local distributor or the closest ECLERIS office, prior to its use. 

5.  OPERATION 

– FUNCTIONS 

This section thoroughly describes the  device functioning.  

This system must be used only for medical, paramedical or health professionals. 

 

1)  Connect the power supply cable to the equipment and the electrical network.  
2)  Turn the EVERLUX on by pressing the general on/off switch. 
3)  Turn the LED system on by turning the light intensity control (dimmer) clockwise. 
4)  Use this control to regulate the light intensity.  

 

The EVERLUX light source has an over-temperature protection. LED system will turn off if 

it reaches a risky operation temperature. 

5)  Once the procedure is over, turn the EVERLUX off in order to maximize the LED system 

lifetime and energy saving. 

6)  Should  you  use  a  video  camera  (Ecleris  PROCAM),  connect  it  to  the  video  input 

connector. 

7)  Connect the video output (rear panel BNC or S-Video connector) to the video input of the 

viewing device and turn the EVERLUX on. 

8)  Select the video input and video Standard in the viewing device. 

6.  MAINTENANCE 

In  order  to  keep  an  optimum  operation  of  the  EVERLUX  it  is  recommended  to  perform  the 
following procedures: 

  Periodic cleaning of all exterior components. 

  Connections inspection. 

  Annual technical inspection of the equipment.  

Summary of Contents for EVERLUX

Page 1: ...ut previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified...

Page 2: ...3 2 INTENDED USER 5 3 3 FEATURES AND DESCRIPTION 5 3 3 1 FRONT PANEL 5 3 3 2 REAR PANEL 5 3 4 COMPONENTS 6 4 INSTALLATION AND SET UP 6 4 1 REVISION 7 5 OPERATION FUNCTIONS 7 6 MAINTENANCE 7 6 1 CLEAN...

Page 3: ...recautions Please read the manual thoroughly before operating the equipment Should you have any doubt regarding the contents of the manual herein or the equipment operation and safety please contact y...

Page 4: ...blem the system must be laid a plane surface and power cable disconnected The device must be used only by personnel informed of all this warnings Any use outside of the manufacturer specific intent us...

Page 5: ...its use for all types of procedures The EVERLUX allows the connection of the Ecleris PROCAM universal endocamera which has a 1 3 CCD sensor and integrated controls especially designed for medical appl...

Page 6: ...l match perfectly and not exceed the adapter 5 In case of using the EVERLUX as a Microscope Colposcope light source mount the EVERLUX on the Ecleris Microscope Colposcope forearm See Light Source Inst...

Page 7: ...S office prior to its use 5 OPERATION FUNCTIONS This section thoroughly describes the device functioning This system must be used only for medical paramedical or health professionals 1 Connect the pow...

Page 8: ...following the instruction provided by their manufacturer Do not turn the light source on until an hour 1 has passed from the cleaning since the unit can get damaged It is recommended to clean the exte...

Page 9: ...UT Composite video CVBS 1 Vpp 75 BNC S Video Y C Jack mini DIN ECLERIS ENDODIGI CONNECTOR Jack 3 5mm mono DIMENSIONS 120x300x155mm width x depth x high MASS 7 7 pounds SERVICE CONDITIONS 50 104 F 35 7...

Page 10: ...Lateral ventilation grids are obstructed Clear the lateral ventilation grids Fan is not working Turn the equipment off immediately and contact the closest Technical Service Light is correct but no im...

Page 11: ...blem description The shipment should be done in the original packaging to avoid damage to the equipment during the transportation ECLERIS will not be responsible for damage caused during the transport...

Reviews: