Press the dose button
until the dose counter shows “0”.
The “0” must line up with the dose pointer.
• Check that insulin squirts from the needle tip.
• Check the insulin window to see that there is no gap
between the black piston rod head and the orange piston,
see illustration 3A.
If no insulin appears
or if there is a gap in the insulin
window,
see illustration 3B,
repeat the insulin flow test until
insulin squirts from the needle tip. If no insulin appears after
several attempts, change the needle and test the insulin
flow again.
The insulin flow test is complete when you see insulin squirt
from the needle tip and when there is no longer a gap
between the piston rod head and the piston.
9
3
Do not try to inject if no insulin appears.
You may
not receive any insulin, even though the dose counter
may move. This could lead to hyperglycemia.
If insulin does not appear after testing the flow
repeatedly,
do not use the pen and call Novo Nordisk at
1-800-727-6500.
3B
3A
8-4251-31-001-X_v1-27:Layout 1 2013-06-28 10:43 AM Page 9