background image

12 

 

SPECIFICATIONS

 

 

RANGE 

45 to 125 kV 

– Model 815 

40 to 115 kV 

– Model 815L 

 

KVP ACCURACY 

2% +/- 1 kV, at 25 to 100 mA, 

18 to 42 cm from head, 

Tungsten target x-ray tube with 1.5 mm 

Al equivalent filtration 

 

EXPOSURE TIME ACCURACY 

1% +/- 2 millisecond - 120 mS and 

longer at 50 to 80 kV, 25 to 50 mA 

Minimum exposure time is 100 

millisecond for measurement 

 

MINIMUM EXPOSURE TIME 

40 millisec - High Speed Mode 
80 millisec - High Resolution Mode 

 

MINIMUM CURRENT (mA) 

7 mA at 65 kV, 10 cm from x-ray (815) 
5 mA at 50 kV, 10 cm from x-ray (815L) 
Also depends on focus 

 

DISPLAY 

0.4" (10.2mm) Liquid Crystal 
8 Character Alphanumeric 

 

CONTROLS/INDICATORS 

ON/OFF Switch 

– Illuminated 

MODE Switch 

– momentary push-button 

 

 

OUTPUTS 

8 Character Liquid Crystal Display 

 

POWER 

4 AA batteries accessible from 

bottom of case 

Low battery indicator 

 

BATTERY LIFE 

48 hours continuous 
Typically 9 months of normal use 

 

ENVIRONMENTAL 

Temperature 

Ope10oC to 40oC 

Storage 0oC to 50oC 

Humidity 

Up to 75% relative humidity 

 

PHYSICAL SIZE 

150 X 120 X 58.5 mm 
5.9 X 4.7 X 2.3 inches 

 

WEIGHT 

0.7 kg, (1.5 lb) 

 

CONNECTIONS 

None 

 
 
 
 
 
 
 

 

OPTIONAL ACCESSORIES 

Carrying Case 815CC 
Hard Carry Case 815HS 
Custom designed case for ECC kVp Meter 

Summary of Contents for 815

Page 1: ...r Exposure Time Meter Rev November 21 2013 Electronic Control Concepts PO Box 182 1160 US Route 50 Milford Ohio 45150 9705 USA Phone 800 VIP XRAY Fax 845 247 9028 sales eccxray com www eccxray com A D...

Page 2: ...AL TABLE OF CONTENTS Page DESCRIPTION 3 GENERAL INSTRUCTIONS SWITCH SETTINGS 4 OPERATION 5 FILTRATION CORRECTION 7 CONNECTIONS 7 BLUETOOTH REMOTE MONITOR 8 STATUS MESSAGES 10 BATTERY REPLACEMENT 10 WA...

Page 3: ...e operator can also determine the characteristics of an x ray After a reading is obtained pushing the Mode button will give the operator a message indicating if the x ray is 1 half wave or 2 full wave...

Page 4: ...tion Mode the delay is approximately doubled When using the instrument in High Speed Mode the x ray exposure time must set to at least 40 ms 0 04 sec When using the instrument in High Resolution Mode...

Page 5: ...sure Time There are two modes requiring different exposure times In the Normal or High Speed Mode the exposure time must be at least 0 04 sec 40 ms Longer exposure times than 50 or 100 ms are not nece...

Page 6: ...that will be added to the exposure time We recommend making exposure time readings at 50 to 80 kV and 25 to 50 mA Also the accuracy specifications apply for measurements greater than 150 milliseconds...

Page 7: ...7 1 7 3 4 For example if your x ray has 2 5 mm Al that is equivalent to 1 0 mm of added filtration A reading at approximately 60 kVp 50 mA will be 1 1 kV high Subtract 1 1 kVp to get the correct read...

Page 8: ...automatically Running the App When the ECC Remote Monitor app is started it asks for your permission to use the Bluetooth capabilities of your Android device if you deny this request the application w...

Page 9: ...count selection menu appears If you choose to upload the data a text file with all the information collected will be stored in your Google Drive cloud storage If you saved your file to the SD card use...

Page 10: ...ease the distance from the x ray head to the Model 815 OvrScale Over Scale The output of the x ray exceeds the measurable range of the Model 815 Low Batt Battery is low The unit may still be operated...

Page 11: ...and workmanship for a period of 2 years There is no warranty on the battery ECC will replace or repair any Model 815 during the first year after shipment that does not show obvious signs of abuse Cont...

Page 12: ...at 50 kV 10 cm from x ray 815L Also depends on focus DISPLAY 0 4 10 2mm Liquid Crystal 8 Character Alphanumeric CONTROLS INDICATORS ON OFF Switch Illuminated MODE Switch momentary push button OUTPUTS...

Page 13: ...To simplify this discussion please refer to the chart below The chart below shows the maximum kV and mA This instrument can accurately measure at kV mA combinations below the three curves at the part...

Page 14: ...units with a single winding head excite the x ray at a lower voltage before the actual exposure in order to allow the filament to warm up Some of these x rays emit radiation during this pre heat time...

Reviews: