background image

 

BLUETOOTH  REMOTE  MONITOR

 

 

This instrument is equipped with a feature that can be used to show output data 
on an Android smartphone or tablet using the Bluetooth module in every 
instrument. When not using the Bluetooth feature, the unit is used in normal 
fashion. In order to use the Remote Monitor, an application (app) must be 
installed on your Android phone or tablet. 

 

This feature is installed on Model 815 kVp Meters serial number 646 and up and 
Model 820 kVp/mA Meters serial number 329 and above. 

 

This feature is provided as a convenience so that it is not necessary to go into 
the x-ray room to observe the meter after each reading. There is also a file save 
capability which can assist in recording data. 

 

How to download the app:

 

1: Open the Google Play Store 
2: Search for ECC Remote Monitor 
3: Download the App 
4: The app will install automatically 

 

Running the App:

 

When the ECC Remote Monitor app is started, it asks for your permission to use 
the Bluetooth capabilities of your Android device, if you deny this request, the 
application will close. 

 

Connecting to a device:

 

The home screen should now be shown, and you may choose to connect to a 
device via Bluetooth in the action bar at the top or bottom of the screen 
(depending on your device screen size). 
You may choose devices by using the menu button in any Android systems 
below version 
3.0 (Honeycomb). If the ECC device does not appear in the connection menu, 
select the scan button and it will appear in the menu if it is within range. 

 

The device name will either be shown as 

null

, or 

ECC_XXX 

where “XXX” is the 

serial number of the device. 
Select the ECC device to connect with it. 

 

The first time you attempt to connect with this device, you will be asked for a 
pairing code, 

which will be „1234‟. A confirmation message will be displayed if the 

device has been successfully connected. You may now take an x-ray or mA 
measurement. 

Summary of Contents for 815

Page 1: ...r Exposure Time Meter Rev November 21 2013 Electronic Control Concepts PO Box 182 1160 US Route 50 Milford Ohio 45150 9705 USA Phone 800 VIP XRAY Fax 845 247 9028 sales eccxray com www eccxray com A D...

Page 2: ...AL TABLE OF CONTENTS Page DESCRIPTION 3 GENERAL INSTRUCTIONS SWITCH SETTINGS 4 OPERATION 5 FILTRATION CORRECTION 7 CONNECTIONS 7 BLUETOOTH REMOTE MONITOR 8 STATUS MESSAGES 10 BATTERY REPLACEMENT 10 WA...

Page 3: ...e operator can also determine the characteristics of an x ray After a reading is obtained pushing the Mode button will give the operator a message indicating if the x ray is 1 half wave or 2 full wave...

Page 4: ...tion Mode the delay is approximately doubled When using the instrument in High Speed Mode the x ray exposure time must set to at least 40 ms 0 04 sec When using the instrument in High Resolution Mode...

Page 5: ...sure Time There are two modes requiring different exposure times In the Normal or High Speed Mode the exposure time must be at least 0 04 sec 40 ms Longer exposure times than 50 or 100 ms are not nece...

Page 6: ...that will be added to the exposure time We recommend making exposure time readings at 50 to 80 kV and 25 to 50 mA Also the accuracy specifications apply for measurements greater than 150 milliseconds...

Page 7: ...7 1 7 3 4 For example if your x ray has 2 5 mm Al that is equivalent to 1 0 mm of added filtration A reading at approximately 60 kVp 50 mA will be 1 1 kV high Subtract 1 1 kVp to get the correct read...

Page 8: ...automatically Running the App When the ECC Remote Monitor app is started it asks for your permission to use the Bluetooth capabilities of your Android device if you deny this request the application w...

Page 9: ...count selection menu appears If you choose to upload the data a text file with all the information collected will be stored in your Google Drive cloud storage If you saved your file to the SD card use...

Page 10: ...ease the distance from the x ray head to the Model 815 OvrScale Over Scale The output of the x ray exceeds the measurable range of the Model 815 Low Batt Battery is low The unit may still be operated...

Page 11: ...and workmanship for a period of 2 years There is no warranty on the battery ECC will replace or repair any Model 815 during the first year after shipment that does not show obvious signs of abuse Cont...

Page 12: ...at 50 kV 10 cm from x ray 815L Also depends on focus DISPLAY 0 4 10 2mm Liquid Crystal 8 Character Alphanumeric CONTROLS INDICATORS ON OFF Switch Illuminated MODE Switch momentary push button OUTPUTS...

Page 13: ...To simplify this discussion please refer to the chart below The chart below shows the maximum kV and mA This instrument can accurately measure at kV mA combinations below the three curves at the part...

Page 14: ...units with a single winding head excite the x ray at a lower voltage before the actual exposure in order to allow the filament to warm up Some of these x rays emit radiation during this pre heat time...

Reviews: