XS series operating manual
XS user manual v0.55
- 94 -
Description of the input output signals
COM_IN
: electrical common between input signals.
COM_OUT
: electrical common between output signals.
CTRLIN (input)
: request for remote control of the measurement unit.
TYPE (input)
: selection of the measurement function (DXS models only)
Logical state HIGH : Megohmmeter.
Logical state LOW : Strength Test.
MES_DCH (input)
: selection of the measurement or discharge state :
Logical state HIGH : Measurement
Logical state LOW : Discharge
N0,N1,N2,N3 (input)
: binary coded data lines for the number of the parameters set :
N3
N2
N1
N0
Memory
0
0
0
0
0
0
0
0
1
1
0
0
1
0
2
0
0
1
1
3
0
1
0
0
4
0
1
0
1
5
0
1
1
0
6
0
1
1
1
7
0
1
0
0
8
1
0
0
1
9
CTRLOUT (output)
: acknowledgment of the CTRLIN signal.
EOT (output)
: the contact of the relay is closed at the end of the test.
PLCPASS (output)
: the contact of the relay is closed when the test pass.
PLCFAIL (output)
: the contact of the relay is closed when the test fail.
ERROR (output)
: the contact of the relay is closed when an error occurs:
‘CHARGE', 'OVERLOAD', 'OVERVOLTAGE', 'BOARD NOT READY',
'SYNCHRO ERROR', 'INTERL
OCK DISABLE', 'VOLTAGE ERROR'…