XS series operating manual
XS user manual v0.55
- 124 -
OHM (OHMmeter) :
Sets the main thresholds unit (Ohm) and the ground continuity display function.
Caution : The action on this command will erase all previous recorded thresholds
(Reset to zero).
VOLT (VOLTmeter) :
Sets the main thresholds unit (Volt) and the ground continuity display function.
Caution : The action on this command will erase all previous recorded thresholds
(Reset to zero).
HTIM(HTIMe) <NR1> :
Sets a new value in second to the current hold time parameter to the parameter set of the
running function. Minimum time for the leakage current function : 2 Seconds
For instance : HTIM 3
RTIM(RTIMe) <NR1> :
Sets a new value in second to the current rise time parameter to the parameter set of the
running function
For instance : RTIM 10
FTIM(FTIMe) <NR1> :
Sets a new value to the current fall time parameter to the parameter set of the running
function.
For instance : FTIM 5
HLIM(HLIMit) <NR3> :
Sets a new value to the high limit parameter to the parameter set of the running function.
For instance :
MEG:HLIM 2.0E+6 (new value in Ohms)
HIP:HLIM 1.45E-4 (new value in Amps)
GND:HLIM 1.00E-1 (new value in milliohm)
The unity is automatic, given in Volt, Ohm or Amp and function dependent.
CBHLIM <NR3> :
Same command as HLIM, but reset the low limit parameter before to change the high limit
value.
LLIM(LLIMit) <NR3> :
Sets a new value to the low limit parameter to the parameter set of the running function.
For instance :
MEG:LLIM 1.0E+6 (new value in Ohms)
HIP:LLIM 3.50E-6 (new value in Amps)
GND:LLIM 5.02E-2 (new value in milliohm)
The unity is automatic, given in Volt, Ohm or Amp and function dependent.
(Aa, Bb, Cc, Dd, Ee, Ff, Gg, Hh) :
Parameters setting a sequence of 8 tests of the selected memory only for SXS models.
Each test is defined by 2 letters : one for the function (M for megohmmeter, H (HIPOT) for
strength, G (GROUND) for earth continuity, X(points=) for test number in case of multiple
earth continuity test, P(Pause) to place a waiting phase between two consecutive tests and
L (LEAKAGE) for leakage current), and one digit to indicate the memory number (cases M,
H, G, L, V). V0 is an empty test. The case X is special, because it can be followed by one or
2 digits. These one indicate the step number of the multiple continuity. This number can be
set from 2 to 99.
Caution : All 8 tests must be carried out (Enter V0 for an empty test).
For instance : SEQ : PAR 0 : (G1,M2,H1,M3,L1,V0,V0,V0)