EasyStand

  

Glider Assembly

FIG. 4

FIG. 1

FIG . 1- Attach the front caster frame to the front frame with 

bolts and tighten securely. Laying your unit on the floor side 

ways may help with lining up the holes. 

FIG . 2- Attach the casters to the rear base frame. Tighten 

securely using the provided wrench. Ensure the casters are 

flush with the rear base frame. The casters are supplied with 

a locking compound pre-applied to the threaded stem. Do not 

remove this material, it is important to prevent the casters 

from loosening over time. 

FIG . 3- Place the front frame and the seat assembly on its 

side, slide the front frame into the opening of the seat 

assembly as shown. Line up holes and insert all four bolts 

first, then place nuts onto bolts and tighten securely with hex 

wrench and 19mm open-ended wrench. Place unit upright on 

its wheels. 

FIG . 4- Remove the ratchet handles and bolts from the glide 

handles, slide the glide handles into the slots. Tighten 

securely. If any additional options were ordered, please refer 

to the drawings on pp. 5-10. Install the seat onto the seat 

assembly after all other options are added.

Front Frame

Seat Assembly

®

®

Videos on EasyStand assembly & fitting can be viewed at easystand.com/videos

Page 3 - Assembly

Transferring - Page 20

For certain individuals, transferring can be a difficult task, especially at first. Like many 

things, transferring gets easier with practice. With assisted transfers, it is always a good 

idea to review the transfer strategy with the client before transferring, and talk them 

through the steps as they are happening. This will help reduce anxiety for the person being 

transferred. There are many different methods of transferring into the EasyStand Glider, 

whether the transfer is independent or with assistance. There are also a variety of features 

on the Glider to help make transferring more manageable. For proper transfer techniques, 

please seek professional advice from a therapist or physician.

Caution: Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider 

before transferring.

Transfer Methods 

Independent Transfer: For those individuals who are able to 

transfer independently, Altimate Medical has designed the Glider 

base so a wheelchair can be positioned at any angle from parallel 

to 90 degrees against the edge of the Glider seat (the front 

wheelchair caster can even come underneath the front corner 

of the seat). The knee pads flip up and out of the way to give the 

person room to bring their legs in and position their feet. The 

knee pad does not lock in the upright position when transferring, 

do not grab the knee pad for support. When transferring out of 

the EasyStand, pump the seat up so it is equal to or higher than 

the seat the user is transferring into.

Stand Pivot Transfer: A caregiver can assist a patient into the 

stander using a stand pivot transfer. The caregiver helping with 

the stand pivot transfer assists the person out of their 

wheelchair to standing, then turns or pivots and sits the person 

down on the seat of the stander.

Patient Lift Transfer: Caregivers can assist people who have a 

higher level of disability by using a patient lift device, or an 

overhead ceiling lift to transfer into the EasyStand. The rear legs 

of the EasyStand Glider are designed to allow room for the 

lifting device to be used from many different angles. To make the 

transfer from a patient lift device, recline the back of the 

standing frame slightly. With some patient lift devices it may 

make the transfer easier to lift the Glider foot plates up to a 

higher position, allowing the transfer device to slide closer to the 

base of the Glider. Be sure to reposition the foot plates and back 

prior to standing.

Transferring

!

FIG. 3

FIG. 2

Use FIG . 1 - FIG . 3 for International orders only .

Summary of Contents for Glider Medium

Page 1: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Page 2: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Page 3: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Page 4: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Page 5: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Page 6: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Page 7: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Page 8: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Page 9: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Page 10: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Page 11: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Page 12: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Page 13: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Page 14: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Page 15: ...I s Customer Service Department Charging the Battery The EasyStand Pow r Up Lift uses two 12 volt sealed lead batteries rated at 2 9 amps each An EasyStand can be lifted approximately 100 times on one full battery charge An audible alarm will sound when the battery is low Please charge the battery after the alarm sounds A battery needing to be fully recharged will take approximately 8 hours The Ea...

Page 16: ...should assure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the rela tive humidity should be at least 30 Mains power quality should be that of a typical com mercial or hospital environment IEC 60601 Test Level Elec...

Page 17: ...lectromagnetic Immunity The EasyStand Glider is intended for use in the electromagnetic environment specified below The customer or the user of the EasyStand Glider should assure that it is used in such an environment Immunity Test Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms IEC 60601 Test Level Electromagnetic Environment Guidance NOTE 1 At 80 MHz and 800 MHz the higher frequency r...

Page 18: ...y not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in wa...

Page 19: ...ion change FIG 14 To adjust the back angle use the knob located on the following arm Before standing the back angle must be set so the indicator on the side of the unit is in the green FIG 15 The back seat angle locator is located on the side of the unit When the unit is in the seated position the green area indicates the back angle Before standing the indicator must be in the green Caution Before...

Page 20: ...G50240 Head Support Tall 25 34 H PNG50037 If you ordered the high mount chest vest bracket see p 9 FIG A Place head support bracket onto accessories mounting bracket as shown Insert 4 bolts and nuts and tighten securely FIG B Insert the head support onto the unit and tighten securely FIG A FIG B Page 17 Options Part Detail Option Assembly Page 6 Glider Options Parts Detail ...

Page 21: ...2 PNG50066 PNG50067 PNG50068 To assemble the back option remove the 2 bolts from the back slide the back assembly into the seat assembly as shown Tighten securely with bolts provided then add the swing away arms Remove the nut and lock washer from the arm Place the arm onto the unit keeping the arm assembly together Place nut and lock washer onto bolt on the inside of the unit Tighten securely to ...

Page 22: ...ng the Glider keep hands clear of the resistance cylinders upper and lower mounting points FIG 5 FIG 7 Adjustments FIG 8 FIG 6 Page 19 Perfect Fit Guide Adjustments Page 4 Perfect Fit Guide step one measure Before transferring into the Glider measure for approximate A seat depth and seat to B foot plate settings Caution These adjustments are only to be made while the unit is unoccupied step two se...

Page 23: ... of features on the Glider to help make transferring more manageable For proper transfer techniques please seek professional advice from a therapist or physician Caution Always lock wheel locks on a wheelchair and the casters on the EasyStand Glider before transferring Transfer Methods Independent Transfer For those individuals who are able to transfer independently Altimate Medical has designed t...

Page 24: ... are specifically designed to be used with Altimate Medical accessories and options Unless otherwise noted accessories and options from other manufacturers have not been tested by Altimate Medical and are not recommended for use with Altimate Medical products Product Modifications made without express written consent including but not limited to modification through the use of unauthorized parts o...

Page 25: ...for children Do not use bleach or any other bleach based cleaners on the EasyStand Disinfecting wipes are safe to use on metal and non porous surfaces Use a damp sponge with mild soap or a mild non abrasive disinfectant Spot clean Neoprene with mild soap and water Neoprene covers may be removed and hand washed with water Air dry the covers Always use care when removing and replacing the covers Do ...

Page 26: ...r back the unit may not descend properly Adjust the back angle toward 90 degrees so it will release properly My hydraulic does not work smoothly it is not leaking oil but acting inconsistent With the unit unoccupied pump it all the way to standing Once the seat is in the standing position pump 12 to 15 additional strokes to bleed any air out and circulate fluid Pull back on the pump handle to retu...

Page 27: ...cal satisfaction to be defective and within warranty period shall be repaired or replaced at Altimate Medical s option ALTIMATE MEDICAL SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR AND OR REPLACEMENT This warranty does not include any labor charges incurred in replacement parts installation Freight charges to factory are at the expense of consumer t...

Page 28: ...rves the right to change features specifications and prices without prior notification Check with Altimate Medical for latest information FORM MANGLID 120115 REV A Copyright 2015 Altimate Medical Inc All rights reserved Printed in the U S A Glider medium and large 262 West 1st St Morton MN 56270 USA Toll Free 800 342 8968 or 507 697 6393 Fax Free 877 342 8968 or 507 697 6900 email info easystand c...

Reviews: