![EasyStand Evolv Owner'S Manual Download Page 16](http://html.mh-extra.com/html/easystand/evolv/evolv_owners-manual_524082016.webp)
page 13
page 26
Pow’r Up Lift Option
Charging the Battery
The EasyStand Evolv Pow’r Up Lift uses two 12 volt sealed lead batteries rated at 2.9 amps each.
An EasyStand Evolv can be lifted approximately 100 times on one full battery charge. An audible
alarm will sound when the battery is low. Please charge the battery after the alarm sounds. A bat-
tery needing to be fully recharged will take approximately 8 hours.
The EasyStand Evolv can be raised to the standing position while
being charged. The function of the charger is to detect a full bat-
tery and then provide a constant potential power supply to hold the
battery at a full condition without overcharging. When the battery
EMF rises to a factory preset point, the charger circuit will not allow
the peak output voltage to exceed that value. This factory setting
protects a battery from excessive depletion of electrolyte, which
can occur from overcharging. The charger can be found in the tool
pouch on the back of the unit.
CAUTION: Use only the Linak charger, model numbers 00CH01-XX
OR 10CH01S-00, provided by Altimate Medical with the Pow’r Up
option.
1. Plug the receptacle into a wall outlet. The light on the charger will
be green indicating an open circuit.
2. FIG . A- Plug the charger cord into the side of the hand control.
The light on the charger will turn yellow indicating the batteries are
being charged.
3. FIG . B- The unit will arrive unplugged from the battery, plug in the
white cord that supplies power to the actuator and the black cord
that supplies power to the remote. The black cord has a notch on it
to orient the connector properly. Fully charge battery before use.
Note: The white cord has a rubber washer on it, which makes it very
sticky and sometimes will not get pushed in completely, make sure
to push the cord all the way into the battery.
4. When the batteries are charged, the light on the charger will turn
green.
5. Fully charge the batteries every three months or less during periods of intermittent or non-use.
CAUTION: Ensure the cables are attached properly per the instructions for use. Do not allow the
cables to drag on the floor or to catch on objects.
FIG . A
FIG . B
!
!
Pow’r Up Lift
The hand pendant
is a Type BF applied
part.