17
3.3.2 Nominal Resistance Deviation Percentage
In percentage display mode, users can set nominal resistance values
to display the deviation percentage between the resistance and the
nominal value, and can also set the upper and lower percentage limits
within the sorting function. By default, configuration starts from the least
significant bit, with the most significant bit reserved for symbol display.
Positive values are displayed as " " ("_" in settings) and negative values
are displayed as "-". These values are used to determine whether the
deviation percentage of the measured resistance from the nominal
resistance exceeds the preset upper and lower percentage limits. The
formula for deviation calculation is:
Measured Resistance − Nominal Resistance
Nominal Resistance
∗
100%
Note: This function is only accessible in percentage display mode,
and the upper percentage limit must be greater than the lower percentage
limit.
3.3.3 Sorting Steps