16
8 Specifications
Specifications are based on a one year calibration cycle and apply from +18
℃
to
+28
℃
unless stated otherwise. All specifications assume a 10 minute warmup period.
8.1 DC Voltage Measurement
Range
Maximum
measurement range
Resolution
Accuracy (% of r Counts)
ETX-1814
ETX-2014
30V
0V
~
31V
0.001V
0.05+2
0.02+2
100mV
-15mV
~
80mV
0.001mV
0.05+20
0.02+20
80mV
~
125mV
0.01mV
0.05+2
0.02+2
-10
℃~
18
℃,
+28
℃~
55
℃
temperature coefficient,±0.005%FS/
℃
.
Input resistance:
>
1MΩ.
8.2 DC mV Output
Range
Maximum output
range
Resolution
Accuracy (% of r Counts)
ETX-1814
ETX-2014
100mV
-15mV
~
99.999mV
0.001mV
0.05+20
0.02+20
100mV
~
125mV
0.01mV
0.05+2
0.02+2
-10
℃~
18
℃,
+28
℃~
55
℃
temperature coefficient,±0.005%FS/
℃
.
Maximum load: 1mA or 1kΩ
(
It should be based on the lower load.
)
Summary of Contents for ET-2714
Page 3: ...10 Warning 19...