14
on 25% increment.
Press
or
instantly to make output jump within the span of 0% to
100%.
6.2 Auto Ramping the Output
Automatic ramping gives you the ability to continuously apply a varying stimulus from
the calibrator to a transmitter, while your hands remain free to test the response of the
transmitter.
When you press
, the calibrator produces a continuously repeating 0 % - 100 % -
0 % ramp in your choice of three ramp waveforms:
0%-100%-0%
40-second smooth ramp
0%-100%-0%
15-second smooth ramp
0%-100%-0%
Stair-step ramp in 25 % steps, pausing 5 seconds at each
step.
Exit the slop output function, please press any key.
6.3 Factory Reset
Factory reset consists of the following items:
The working status recovers to the voltage measurement.
Automatic shutdown time is reset to be 30 min, which is effective.
LCD backlight brightness is set to be 60%.
Output span is recovered to be factory default.
Start the calibrator and press
until the buzzer works, and the recovery of factory
default will enter working mode when the recovery is completed.