Table 10: System Compliance Information
–
Enclosure Port Immunity to RF Wireless
communications equipment
Test Frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
power (W)
Distance(m)
Immunity
Level
(V/m)
Compliance
Level
(V/m)
385
380
–
390
TETRA 400
Pulse
modulation b)
18 Hz
1,8
0,3
27
27
450
430
–
470
GMRS 460,
FRS 460
FM c)
± 5 kHz deviation
1 kHz sine
2
0,3
28
28
710
704
–
787
LTE Band 13, 17
Pulse modulation b)
217 Hz
0,2
0,3
9
9
745
780
810
800
–
960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2
0,3
28
28
870
930
1 720
1 700
–
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,
3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2
0,3
28
28
1 845
1 970
2 450
2 400
–
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450, LTE
Band 7
Pulse
modulation b)
217 Hz
2
0,3
28
28
5 240
5 100
–
5 800
WLAN 802.11
a/n
Pulse
modulation b)
217 Hz
0,2
0,3
9
9
5 500
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
Summary of Contents for InSight AME-01350-US
Page 1: ......
Page 2: ...EarlySense Modular Bedside Unit User Guide ALL IN ONE SAFETY VITALS EarlySense InSightTM 0344...
Page 15: ...14 Blank page for double sided printing...
Page 24: ...23 Blank page for double sided printing...
Page 35: ...Appendix 34 B Appendix FCC Compliance Standard Statement...