21
Operating Conditions
Table 5: Operating Conditions
Relative Humidity
Sensing Unit 30%
−
95% Non-condensing
InSight Device 15%
−
93% Non-condensing
Ambient Operating Temperature
Sensing Unit and InSight Device: 5
−
40
C
(41
−
104
F)
Atmospheric Pressure
InSight Device 700
−
1,060 hPa
Sensing Device 5
00−1,060 hPa
Storage and Transportation Conditions
Table 6: Storage and Transportation Conditions
Temperature Range
-20
−
50
C (-4
−
122
F)
Relative Humidity
InSight Device:15%
−
90% Non-condensing
Sensing units:10%-90% Non -condensing
Atmospheric Pressure
InSight
Device 700−1,060 hPa
Sensing Device 500−1,060 hPa
Cables Used with the Sensing Units and EarlySense InSight Device
Table 7: Cables
Connector
Name
Connector Type
Type of Cable
Length
(m)
# of Identical
Connectors
Sensing Unit
Extension Cable Sensor Extension Cable
Unshielded
≤
12
2
LAN
RJ45
Shielded
>3
1
1.4
Additional Specifications
1.4.1
System Performance
–
Sensing Unit
Table 8: Performance Specifications
Respiratory Rate
Heart Rate
Movement
Range
6
–
45 Br./min.
30
−
170 BPM
0, L, M, H, EH
Averaging Period
1 Min.
1 Min.
15 Seconds
Accuracy
±4% or ±1.5 Br./min.,
whichever is greater
±4% or ±5 BPM, whichever is greater
±4% or ±1.5 Br./min.,
whichever is greater
±4% or ±5 BPM, whichever is greater