background image

No Fault: Solid green

Level: Solid red

Temp: Off

4. To change the test mode, press and hold the RS button until the No Fault LED changes state 

(about 10 seconds). If in NiCad mode, the LED will be solid green. If in lead-acid mode, the 
LED will start flashing. 

4.3 Verify Sensor Voltage

 

While unlikely, it's possible the ELM sensor isn't receiving the required voltage for normal 
operation. Each sensor has a minimum working voltage of 5.5 volts DC.

1. Remove the sensor cover. 

2. Using a meter, measure the voltage at the points shown below. Note that it doesn't matter 

which side the ribbon cable is connected to, the measurement points are the same. 

3. If the voltage is under 5.5VDC, replace the ribbon cable. If the voltage is still low, it could be 

the Vigilant sensor. 

4.4 Move ELM Sensor

 

The most unlikely scenario for erroneous ELM alarms could be interference from something inside 
the battery jar. This can be challenging to troubleshoot because it could only affect a handful of 
ELM sensors on the same battery. One example of this could be an internal post that is reflecting 
light back at the ELM infrared sensors. 

The only way to troubleshoot this issue is if something is visibly in line of the ELM sensor, try 
moving the sensor to a different spot on the battery. In some scenarios, this can resolve false 
alarms on cells that are problematic. 

Summary of Contents for Vigilant ELM Series

Page 1: ...least likely causes 2 0 How Level Alarms Work Before troubleshooting the ELM it s important to understand how alarming works The ELM is independent from the Vigilant system in that each sensor has fu...

Page 2: ...re the Vigilant has a chance to poll it 3 times that alarm will never register in the software 3 0 Clearing ELM Level Alarms This section will outline how to clear an ELM alarm when the level is confi...

Page 3: ...nce of user error These steps will assume you understand how to login to the Vigilant 1 Login to the web interface and open the Settings section 2 If using version 1 3 11 or later click the ELM Manage...

Page 4: ...age click the dropdown menu and select Calibrate 8 Under the list of sensors select the sensor that needs calibration Alternatively select sensor 1 hold the shift button and then select the last senso...

Page 5: ...ed by qualified personnel If a sensor is installed incorrectly it can throw false alarms When a sensor is mounted using the ELM cradle it is aligned on the cell using the 2 pointed arrows on the sides...

Page 6: ...nfirm Proper Measurement Mode The ELM has two measurement modes lead acid and nickel cadmium NiCad From the factory all ELM sensors ship in lead acid mode If the ELM is in the wrong mode it could caus...

Page 7: ...he ribbon cable is connected to the measurement points are the same 3 If the voltage is under 5 5VDC replace the ribbon cable If the voltage is still low it could be the Vigilant sensor 4 4 Move ELM S...

Page 8: ...as ELM Case Temperature It does not report an actual value rather it alarms against a set value on the sensor Below are the settings available 95 F 35 C Default setting from factory 120 F 49 C 145 F 6...

Page 9: ...the RS button will advance the alarm setting by one 2 So if the power LED is flashing once pressing the RS button briefly will advance the program to setting two 120 F Setting two can be advanced to...

Reviews: