Eagle Eye GD-3000 User Manual

 

Selecting Default Configuration – “dEF” 

To select the Default Configuration, in normal mode, press the 

Next 

button twice to get 

to “Con” or the Configuration menu.Then press the 

Enter 

button to enter the Con menu. 

The 

first 

selection is the “dEF” or Default setting. Press 

Enter

. If it is already in Default 

configuration, there will be no action. If it is not already in Default configuration, “nO” will 
be displayed. Press 

Next 

to change it to “YES” (flashing) then press 

Enter 

to confirm 

the change (solid) and press 

Enter 

again to return to “dEF” in the Con menu. Press 

Next 

until “End” is displayed then press 

Enter 

to return to normal operation. 

Selecting Power Up Test Option – “PUt” 

To select the 

Power Up Test 

Configuration, in normal mode, press the 

Next 

button 

twice to get to “Con” or the Configuration menu. Then press the 

Enter 

button to enter 

the Con menu. Then press the 

Next 

button to get to 

the second selection 

“PUt” or 

Power Up Test 

setting. Press 

Enter

. If the test is “On” press 

Next 

to turn it “OFF” 

(flashing) then press 

Enter 

to confirm the change (solid) and press 

Enter 

again to 

return to “PUt” in the Con menu. Press 

Next 

until “End” is displayed then press 

Enter 

to 

return to normal operation. 

Selecting Display Option – “dSP” 

To select the Display Configuration, in normal mode, press the 

Next 

button twice to get 

to “Con” or the Configuration menu.Then press the 

Enter 

button to enter the Con menu. 

Then press the 

Next 

button twice to get to 

the third selection 

“dSP” or Display setting. 

Press 

Enter

. If the display is “On” press 

Next 

to turn it “OFF” (flashing) then press 

Enter 

to confirm the change(solid) and press 

Enter 

again to return to “dSP” in the Con 

menu. Press 

Next 

until “End” is displayed then press 

Enter 

to return to normal 

operation. 

Selecting Buzzer Option – “bUZ” 

To select the Buzzer Configuration, in normal mode, press the 

Next 

button twice to get 

to “Con” or the Configuration menu. Then press the 

Enter 

button to enter the Con 

menu. 

The fourth selection 

is the “bUZ” or Buzzer setting. Press 

Next 

three times to 

get to “bUZ” then press 

Enter

. If the display is “On” press 

Next 

to turn it “OFF” (flashing) 

then press 

Enter 

to confirm the change (solid) and press 

Enter 

again to return to “bUZ” 

in the Con menu. Press 

Next 

until “End” is displayed then press 

Enter 

to return to 

normal operation. 

Selecting Alarm Relay Setting – “ArS” 

To select the Alarm Relay Setting, in normal mode, press the 

Next 

button twice to get to 

“Con” or the Configuration menu. Then press the 

Enter 

button to enter the Con menu. 

13 

Summary of Contents for GD-3000

Page 1: ...Eagle Eye GD 3000 User Manual Updated 10 09 13 4031 W Kiehnau Ave Milwaukee WI 53209 TEL 1 414 962 3377 TOLL FREE 1 877 805 3377 FAX 1 414 962 3660 www eepowersolutions com V1 0...

Page 2: ...ormation 7 3 2 Location 7 3 3 Installation 7 3 4 Power Up 11 3 5 Operation 12 3 6 Default Configuration Factory Settings 12 3 7 Onboard Diagnostics 15 3 8 Sensor Poisons 16 4 MAINTENANCE 16 4 1 Cleani...

Page 3: ...untrained or unqualified persons or use that is not in accordance with these User Instructions may adversely affect product performance and result in sickness or death Use only for monitoring the gas...

Page 4: ...ibration gas at the required concentration level Do not calibrate with expired calibration gas If the instrument cannot be calibrated do not use until the reason can be determined and corrected Do not...

Page 5: ...GD 3000 Combustible Gas Detector uses a microcomputer controlled electronic system to measure the concentration of combustible gas actuate relays and provide a 4 20mA 0 50 LEL output The GD 3000 Combu...

Page 6: ...ectable at 1 2 3 4 5 6 7 8 9 10 default 11 12 13 14 15 16 17 18 19 20 LEL Fan Delay Settings of 0 1 3 default 5 and 10 minutes Fan Minimum Run Time settings are OFF default 3 5 10 or 15 minutes Alarm...

Page 7: ...the GD 3000 Combustible Gas Detector where the normal ambient temperature is below below 0 F or exceeds 125 F 18 C or above 52 C If the target gas is lighter than air natural gas Methane or Hydrogen...

Page 8: ...that the disable setting will cause the alarm relay not to engage at all 7 The dry contact SPDT fan relay has three terminals The common COM normally open N O and the normally closed N C contact The f...

Page 9: ...Gas Detector high on a wall about one foot down from the ceiling If the gas used is propane LP mount the GD 3000 Combustible Gas Detector on a wall or column one foot above the floor 3 See applicable...

Page 10: ...cked up power to Eagle Eye GD 3000 Combustible Gas Detector 3 Eagle Eye GD 3000 Combustible Gas Detector dry contact fan relay can use panel power or fan power Check ratings on panel fan 4 Match fan r...

Page 11: ...ute the test cycle any time power is dropped and reapplied i e power failure During the self test cycle the unit will display the firmware version number then count down from 60 to 0 and finally go in...

Page 12: ...FAn and ALr and sound as outlined above 3 6 Default Configuration Factory Settings The default Display setting is Off The default Power Up Test setting is On The default Buzzer setting is Off The def...

Page 13: ...is displayed then press Enter to return to normal operation Selecting Display Option dSP To select the Display Configuration in normal mode press the Next button twice to get to Con or the Configurati...

Page 14: ...e FrS or Fan Relay setting Press Next six times to get to FrS then press Enter If the fan relay is OFF disabled press Next to change it to 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 or 20 LEL fla...

Page 15: ...d verify unit operations If a problem is found the unit will switch to a fail safe error mode or trouble condition In this error mode the Fan and Alarm relays will be activated the 4 20mA current loop...

Page 16: ...er serviceable parts and substitution of components may impair intrinsic safety which may adversely affect product performance and result in sickness or death CAUTION Avoid the use of harsh cleaning m...

Page 17: ...er to the General section of this User Instruction 6 Locate the switch labeled ENTER TEST on the left side of the printed circuit board Press the Test switch once 7 The Eagle Eye GD 3000 Combustible G...

Page 18: ...or calibration verification test bump test to ensure proper performance of the monitor Failure to do so may adversely affect product performance and result in sickness or death When performing a cali...

Page 19: ...the regulator Ensure that the gas used for calibration matches thegas selected in the Eagle Eye GD 3000 Combustible Gas Detector configuration 3 Assemble regulator hose and test hood and place the tes...

Page 20: ...00 Combustible Gas Detector configuration 1 Check the pressure gauge If there is 25psi or less the cylinder should be replaced 2 Place the test hood over the gas sensor Turn on the regulator to start...

Page 21: ...vicing return unit to factory for servicing d Detector has 4 20mA option set to OFF Set 4 20mA option to On and repeat the test 6 Remove the gas from the sensor Re assemble the Eagle Eye GD 3000 Combu...

Page 22: ...may be present and could result in overexposure and cause sickness or death 4 3 1 Zero the Sensor 1 Remove the Philips screw on the front of the Eagle Eye GD 3000 Combustible Gas Detector Pull the fro...

Page 23: ...and the calibration will progress The display will show this for a maximum of 165 seconds 10 When the calibration is successful the display will flash back and forth between 10 and PA S 11 Remove the...