© EA Elektro-Automatik, DE-41747 Viersen, Helmholtzstr. 31-33, Tel. 02162-3785-0, Fax. 02162-16230
14
Analogue interface
Pin connection on the analogue interface, Sub-D 15 pole
1 0V-Reference
2
+ Reference (0...10V)
3
UPS (Input preset voltage potentiometer 0...10V
4
UPOT (output preset voltage potentiometer 0...10V
5
IPS (Input pre set current potentiometer 0...10V)
6
IPOT (output pre set current potentiometer 0...10V)
7
IMON (output actual value current 0...10V)
8
UMON (output actual value voltage 0...10V)
9
SB (Standby) control input: +5V = OFF (standby), open = ON
10
GND (GND/0V for external programming)
11
LOCK (safety loop / inhibit for HV load)
12
+5V output (for standby mode connect to pin 9)
13 N.C.
14
+15V output
15 N.C.
Fundamental settings for normal operation
For normal operation the following connections must be carried out on the analogue interface:
Pin 3 - Pin 4,
Pin 5 - Pin 6,
Pin 10 - Pin 11
These connections are already made at the included Sub-D plug.
Output terminals on the rear side
HI
(HV socket)
+ or – output on the central contact, depending on
polarity (standard is +). The screen is connected to the
safety ground PE and may never be connected to the load.
LO
(black)
– output
(yellow)
PE safety earth, return path for the load
Interfaces (according to setup)
24-pole socket
IEEE bus interface (if equipped)
9-pole Sub-D connector
RS-232 interface (if equipped)
15-pole Sub-D connector
Analogue interface (must always be plugged)
RJ45 connector
Ethernet port (if equipped)