10

操作控制件

 / Operating controls

前板

1. 

电源开关

2. 

输出开/关(待机)切换开关

位于

On

 

打开输出电压

位于

Off

 

关闭输出电压

3.

 

复位开关

用于恢复至

OVP

值。

.

4. Preset/Normal OVP

切换开关

开关位于

"Preset"

时,可用螺丝刀转动电位器

 OVP  (No. 5) 

设定所需

OVP

极限值。设定完成后需调回正常操作模式!

5. OVP

极限设定用电位器

该电位器用于设置

OVP

极限电压。

6. Local-

本机开关

该开关可将

IEEE

总线模式调回标准模式。

7. Preset/Actual

电压切换开关

开关位于

"Preset"

时,电压表

13

指示预设电压,用电位器

9

可调节。位于“

Actual

”时,指示实际电压。

8. Preset/Actual

电流切换开关

开关位于

"Preset"

时,电流表

14

指示最大预设输出电流,

用电位器

12

可调节。位于“

Actual

”时,指示实际电流。

9. 

电压调节用电位器

 

该电位器用于调节输出电压。

10. 

电流调节用电位器

 

该电位器用于调节最大输出电流。

11. 

电压表

电压表显示预设电压、实际电压和

OVP

电压。

12. 

电流表

电流表显示预设和实际输出电流。

Front panel 

1. Mains switch

2. Switch output on/off (standby)

Position 

on 

= output voltage on

Position 

off 

= output voltage off

3.

 

Reset switch

Serves to reset the OVP.

4. Switch Preset/Normal OVP

In position "Preset"  the desired OVP threshold can be set 

by means of a screw driver with the trimmer  OVP  (No. 5) 

Switch back to normal operation after setting it!

5. Trimmer for setting OVP threshold

This trimmer sets the OVP threshold voltage.

6. Local

This switch is used to switch from IEEE bus mode back to 

standard mode.

7. Switch Preset/Actual Voltage

In position Preset the voltmeter 13 indicates the preset 

voltage, adjusted with potentiometer 9. In position Actual 

the actual voltage is indicated.

8. Switch Preset/Actual current

In position Preset the amperemeter 14 indicates the preset  

max. output current, adjusted with potentiometer 12. In 

position Actual the actual current is displayed.

9. Potentiometer Voltage 

Potentiometer for the output voltage adjustment.

10. Potentiometer Current

Potentiometer for the max. output current.

11. Voltmeter

The  voltmeter  displays  the  preset,  actual  and  the  OVP 

voltages. 

12. Ammeter

The ammeter displays the preset and the actual value of 

the output current.

Summary of Contents for EA-HV 900 Series

Page 1: ...Laboratory Power Supply Series EA HV 9000 Art Nr 26100103 114...

Page 2: ......

Page 3: ...resistance 1 10Hz 0 2 of RLnom 10 100Hz 1 of RLnom 100 10 000Hz 10 of RLnom 8 Stability over 8 hrs 0 01 Temperature stability 50ppm K Ripple 0 05 CURRENT CONTROL Adjustment range 0 100 Regulation accu...

Page 4: ...rnally set by means of an external voltage of 0 10V for 0 rated value PLC control The two external monitor outputs U I each provide an output voltage of 0 10V for 0 rated value InconnectionwiththeIEEE...

Page 5: ...f the unit Further details about measuring and control features of the IEEE or Ethernet option are available in the option s user instruction manual High voltage output The high voltage output is loca...

Page 6: ...indicated on the voltmeter after activatingthe OVP switchandthe Preset LEDlightson If the output voltage becomes for any reason higher than the preset voltage e g operators fault defective compo nent...

Page 7: ...nding The unit may only be operated using a proper ly wired and grounded mains plug as the grounding of the unit leads via the earth line of the power cable This safety feature must not be disabled by...

Page 8: ...n external voltage of zero volt 0V is equivalent to 0Aon the output 10V external is equivalent to the nominal output current of the power supply The external voltage is connected according to the pin...

Page 9: ...ndardized to 10V with an accuracy of 0 2 The monitor signals are available on the respective outputs Pin 7 IMON Pin 8 UMON The end of the monitor cable should be terminated by a resistor e g 100k Ohm...

Page 10: ...Trimmer for setting OVP threshold This trimmer sets the OVP threshold voltage 6 Local This switch is used to switch from IEEE bus mode back to standard mode 7 Switch Preset Actual Voltage In position...

Page 11: ...2 1 Mains switch 2 Output on off standby 3 IEEE Reset resets the IEEE bus 4 Preset Normal OVP Preset Normal adjustment OVP 5 OVP Adjusting the OVP threshold 6 Local only IEEE Switch back to normal mod...

Page 12: ...EA Elektro Automatik DE 41747 Viersen Helmholtzstr 31 33 Tel 02162 3785 0 Fax 02162 16230 12 Rear view IEEE BUS RS 232 9pol Analogue Interface 15 pol Mains Output terminals LO HI...

Page 13: ...3785 0 Fax 02162 16230 13 Sub D 15 1 0V 2 0 10V 3 UPS 0 10V 4 UPOT 0 10V 5 IPS 0 10V 6 IPOT 0 10V 7 IMON 0 10V 8 UMON 0 10V 9 SB 5V offen 10 GND GND 0V 11 LOCK HV 12 5V 9 13 14 15V 15 3 4 5 6 10 11 Su...

Page 14: ...5V output for standby mode connect to pin 9 13 N C 14 15V output 15 N C Fundamental settings for normal operation For normal operation the following connections must be carried out on the analogue in...

Page 15: ...onnel On units above 300V the 0V HV output is internally connected with the safety ground connector PE In case the output has been reversed than the output is connected to the safety ground connector...

Page 16: ......

Page 17: ......

Page 18: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 33 41747 Viersen Germany Tel 49 2162 37 85 0 Fax 49 2162 16 230 ea1974 elektroautomatik cn www elektroautomatik cn...

Reviews: