EA EA-HV 900 Series Manual Download Page 7

© EA Elektro-Automatik, DE-41747 Viersen, Helmholtzstr. 31-33, Tel. 02162-3785-0, Fax. 02162-16230

7

安装

操作产品前,需检查产品外壳、控制件等是否有机械损
坏。

必须使用所供

Sub-D

插头来连接。

如发现有任何损坏,不可将产品连到市电。打开产品前断
开电源插头。

应由受训工程人员执行本机的服务、维修或校正。

本机仅能在铭板规定电压下操作。如需更换保险丝,有一
点极其重要,即:仅能用与原厂所配保险丝规格和外形尺
寸相同的保险丝来替换。

 

更换保险丝时应将本机与市电断开。

接地

本机仅可在接地良好的电源插头下操作。因为它通要过电
源线的地线作为产品引线的地。该安全特征不可因无接地
线的扩展线而中断。

制冷

任何时候都必须保持空气流通顺畅,这点非常重要。

注意!

本产品会产生危险电压。故非受训人员不能操作改产
品,也不可打开产品外壳!

过温保护

 (OT)

如果产品过热(如:风扇坏,进风孔和排气孔太脏等),产
品将自动关闭,且

"OT"

灯会亮。产品冷却后又自动打开。

OT LED

会指示模拟接口的

LOCK

信号没有连接。

环境条件

操作过程中,不论是满载操作还是恒定值操作,环境温度
应在

0...50°C

之间。储存温度可为

-40°C

+70°C

。相对湿

度不应超过

90%

,且无凝露。

Installation

Before taking the unit into operation it is necessary to inspect 

the housing, the controls etc. for signs of physical damage. 

The supplied Sub-D plug must be connected.

If any damage is found, the unit may not be operated on the 

mains. Disconnect the mains plug before opening the unit.

Servicing, repairs or calibrations should only be carried 

out by trained engineers. 

The unit must be operated only on the voltage stipulated 

on the type plate.  If it is necessary to change the fuse, it is 

imperative that it is only replaced by one of same value and 

physical dimensions as the original supplied fuse. 

The unit must be disconnected from the mains whilst 

replacing the fuse.

Grounding

The unit  may only be operated using a proper ly wired and 

grounded mains plug as the grounding of the unit leads 

via the earth line of the power cable. This safety feature 

must not be disabled by using an extension cable without 

a ground lead.

Cooling

It is important that the air circulation remains unimpeded 

at all times.

ATTENTION!

The unit generates hazardous voltages. It must not 

be operated by untrained personnel and not with 

open cover!

Overtemperature protection (OT)

If the unit is overheated (e.g. fan defective, ventilation in- 

and outlets dirty etc.), it will automatically switch off and 

the LED 

"OT"

 will light on. After cooling down the unit will 

switch on automatically.

The OT LED also indicates that the LOCK signal on the 

analog interface (rear side of device) is not bridged.

Ambient conditions

During operation, at full load or constant operation, the 

ambient temperature may lie between 0...50°C. The storage 

temperature can be between -40°C and +70°C. The relative 

humidity should not exceed 90%, non-condensing.

描述

 / Description

Summary of Contents for EA-HV 900 Series

Page 1: ...Laboratory Power Supply Series EA HV 9000 Art Nr 26100103 114...

Page 2: ......

Page 3: ...resistance 1 10Hz 0 2 of RLnom 10 100Hz 1 of RLnom 100 10 000Hz 10 of RLnom 8 Stability over 8 hrs 0 01 Temperature stability 50ppm K Ripple 0 05 CURRENT CONTROL Adjustment range 0 100 Regulation accu...

Page 4: ...rnally set by means of an external voltage of 0 10V for 0 rated value PLC control The two external monitor outputs U I each provide an output voltage of 0 10V for 0 rated value InconnectionwiththeIEEE...

Page 5: ...f the unit Further details about measuring and control features of the IEEE or Ethernet option are available in the option s user instruction manual High voltage output The high voltage output is loca...

Page 6: ...indicated on the voltmeter after activatingthe OVP switchandthe Preset LEDlightson If the output voltage becomes for any reason higher than the preset voltage e g operators fault defective compo nent...

Page 7: ...nding The unit may only be operated using a proper ly wired and grounded mains plug as the grounding of the unit leads via the earth line of the power cable This safety feature must not be disabled by...

Page 8: ...n external voltage of zero volt 0V is equivalent to 0Aon the output 10V external is equivalent to the nominal output current of the power supply The external voltage is connected according to the pin...

Page 9: ...ndardized to 10V with an accuracy of 0 2 The monitor signals are available on the respective outputs Pin 7 IMON Pin 8 UMON The end of the monitor cable should be terminated by a resistor e g 100k Ohm...

Page 10: ...Trimmer for setting OVP threshold This trimmer sets the OVP threshold voltage 6 Local This switch is used to switch from IEEE bus mode back to standard mode 7 Switch Preset Actual Voltage In position...

Page 11: ...2 1 Mains switch 2 Output on off standby 3 IEEE Reset resets the IEEE bus 4 Preset Normal OVP Preset Normal adjustment OVP 5 OVP Adjusting the OVP threshold 6 Local only IEEE Switch back to normal mod...

Page 12: ...EA Elektro Automatik DE 41747 Viersen Helmholtzstr 31 33 Tel 02162 3785 0 Fax 02162 16230 12 Rear view IEEE BUS RS 232 9pol Analogue Interface 15 pol Mains Output terminals LO HI...

Page 13: ...3785 0 Fax 02162 16230 13 Sub D 15 1 0V 2 0 10V 3 UPS 0 10V 4 UPOT 0 10V 5 IPS 0 10V 6 IPOT 0 10V 7 IMON 0 10V 8 UMON 0 10V 9 SB 5V offen 10 GND GND 0V 11 LOCK HV 12 5V 9 13 14 15V 15 3 4 5 6 10 11 Su...

Page 14: ...5V output for standby mode connect to pin 9 13 N C 14 15V output 15 N C Fundamental settings for normal operation For normal operation the following connections must be carried out on the analogue in...

Page 15: ...onnel On units above 300V the 0V HV output is internally connected with the safety ground connector PE In case the output has been reversed than the output is connected to the safety ground connector...

Page 16: ......

Page 17: ......

Page 18: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 33 41747 Viersen Germany Tel 49 2162 37 85 0 Fax 49 2162 16 230 ea1974 elektroautomatik cn www elektroautomatik cn...

Reviews: