background image

14

The Recording  menu allows a  measurement dataset. You can choose between a planned or a continuous dataset.
Memory capacity of the instrument is up to 

12,000

 measurement points or 

50

 datasets.

A continuous dataset can be carried out using  AMI 300 and is composed of several dated measuring points. The 
operator can choose an automatic or a manual dataset, with an instant value or an average. This datasets can't be set 
using Datalogger-10 Software. 

manual dataset 

is composed of measuring points selected by the operator. 

 

a

. Click on 

OK 

or  

 to enter into sub function.

 b

. Select 

Manual 

 with 

 

and

 

. Confirm wih 

OK

.

 c

. Select 

Name 

 with 

 

and

 

. Confirm wih 

OK 

or  

.  Enter dataset name with  arrow keys 

 

  and

 

e

. Confirm wih 

OK

.

 d.

 For measurement launching,  click on 

OK

 with the access key. The number of points selected and the parameter are displayed.

 e.

 To save your dataset  click on 

Save

 with the access key.

1. Create or launch a continuous  dataset 

1.1 Manual dataset

Recording

IV – Menus

1. Connect the hotwire probe to  AMI 300. Probe menu appears.
2. Slide down protection tube.
3.The probe must be perpendicular to airflow : the red point at the bottom of the probe must 
face airflow.
4. Press 

OK 

to enter in the 

Measure 

menu, the air velocity and temperature values are 

displayed.

Using a hotwire

Using  Pitot tube

Dynamic pressure

 = PT - PS

 (-) Static 
pressure PS

(+) Total 
pressure PT

+

-

1. Connect the pitot tube to AMI 300.
2. The probe must be placed at the center of the duct and parallel to the air flow. 
3. Press 

OK 

to enter in the 

Measure 

menu, the air velocity and temperature values are 

displayed on the screen.

An 

automatic dataset 

is composed of measuring points with interval of time.

a

. Click on 

OK 

or  

 to enter  sub function.

b

. Select 

Auto. 

 with 

 

and

 

. Confirm wih 

OK

c

. Select 

Name 

 with 

 

and

 

. Confirm wih 

OK 

or  

.  Enter dataset name with the arrow keys 

 

  and

 

e

Confirm wih 

OK

.

d

. Enter dataset time  and  interval of time between 2 measurement by selecting 

Period

 with access key. Select 

Duration 

 or 

Interval 

with 

 

and

 

. Confirm wih 

OK

. Enter minutes and seconds  with arrow keys

 

and

 

 ( from 1 minutes to 24 hours for the duration and from 

5 seconds to 10 minutes for the interval). Confirm with 

OK

.

e.

 Select 

Start 

for dataset launching.

1.2 Automatic dataset

Tube protection

Red point

 

Sensing element 
(temperature)

Sensing element 

(air velocity)

Airflow 

direction

This sub-function allows to enable or disable the 

key lock

. Click on 

OK 

or  

 to enter into sub function. Select respectively 

ON 

or 

OFF 

with 

 

and

 

   

in order to enable or disable this function. 

Confirm wih 

OK

.  

If the locking is enabled, the code menu appears

This sub-function allows to enter  the 

security code

. Click on 

OK 

or  

 and the code appears. Enter the first digit of the code 

 

with 

 

and

 

  

then move to the next one with 

.

 Confirm wih 

OK

.  

 

Key locking

 

Code

Summary of Contents for AMI 300

Page 1: ...Multifunctional IAQ Monitor AMI 300 P a Supplied with Calibration certificate...

Page 2: ...Airflow 11 Area 11 Duct type 11 Sizes 11 K2 factor 11 Units 11 COmax 11 Air velocity 12 Average 12 Point point average 12 Automatic average 12 Automatic point point average 12 Configuration 12 Thermo...

Page 3: ...ug of thermocouple K J or T type Class 1 as per IEC 584 3 Pressure 2 pressure connectors 6 2 mm made of nickel brass 2 threaded pressure connectors 4 6 mm made of nickel brass for 500 and 200 mbar 1 t...

Page 4: ...0 03 area cm2 1 m3 h m3 h cfm l s m3 s 14 mm VANE PROBE PITOT TUBE DEBIMO BLADES C F ppm ppm RH From 20 to 80 C 0 1 C From 0 to 5 000 ppm From 0 to 1 000 ppm From 5 to 95 RH 1 ppm AIR QUALITY PROBES C...

Page 5: ...r protection Top view Module connections mini Din C1 connector Mini Din C2 connector Graphic display Hour Value Unit Channel Circular menu Memory status used Battery level Keypad On Off key Access key...

Page 6: ...2 channels with current voltage input cables or ammeter clamps 2 Pressure module It allows differential pressure air velocity or airflow measurements with Pitot tube or Debimo on two pressure inputs...

Page 7: ...a connection with right and left keys Connections can be activated or deactivated with or Info Measure Params Select a sub menu with access keys Measurement Communication interrupted Check probes conn...

Page 8: ...Del a RF probe B Select a wireless probe already created B1 Power up the RF probe short press on Multifunction button B2 Go to Probe menu B3 With arrows keys and go to RF probes Display All the RF pr...

Page 9: ...ulb globe temperature For hygrometry probe coupled with black ball thermometer If WBGT index is selected press then OK or and a list appears Select Inside or Outside with arrow and Confirm with OK The...

Page 10: ...F with and in order to enable or disable this function Confirm with OK Units K2 factor IV Menus Sizes Press or OK to enter into sizes sub function You can choose an air vent already registered by sele...

Page 11: ...on beforehand defined You must enter duration click on the Period icon Select minutes or seconds with arrow buttons and Scroll digits with and Confirm with OK The numbers of points is displayed Press...

Page 12: ...allows to enable or disable the screen saving Click on OK or to enter and a list appears Enter the delay before the sreen saving activation with and Confirm wih OK Screen Saving IV Menus This sub fun...

Page 13: ...ng Pitot tube Dynamic pressure PT PS Static pressure PS Total pressure PT 1 Connect the pitot tube to AMI 300 2 The probe must be placed at the center of the duct and parallel to the air flow 3 Press...

Page 14: ...ed dataset is displayed From this screen you can Select other dataset using arrow keys and Dislplay data of other channels using arrow keys and d Click on OK to display selected dataset graph From thi...

Page 15: ...r After Sales Service required for appraisal Maintenance Warranty Battery When battery indicator flashes it is recommended to change the batteries 1 Remove the front part at the back of the instrument...

Reviews: