THERMOPORT 20 english
AutoOff:
AutoOff is used to define an interval in minutes after
which the instrument switches off automatically. The setting OFF
(<1) deactivates the automatic switch off. Select with and ;
confirm with .
One point temperature calibration
CAL1:
CAL1 (one point calibration) gives the user the possibility to
define an offset for temperature 1. The offset is displayed in the
lower part of the screen. Maximum offsets are:
PT100 channel: +/- 2.5°C/°F in 0.01°C/°F steps
TC channel: +/- 10°C/°F in 0.1°C/°F steps.
The offset will be valid over the complete measuring range.
Factory calibration can be obtained by setting the offset to 0.0.