11
3.6.2. Installation of the Sensing Probe into the Gauge Mounting Block
Make sure that the pipe line is depressurized!
Make sure that the pipe line is depressurized!
•
Remove transport protection cap of the head of the sensing probe.
•
Mount the sensing probe in the gauge mounting block in such a way that the alignment pin fits in
the recess on the gauge mounting block.
•
Screw the retainer nut by hand so far that a certain resistance is noticeable.
•
Check the correct installation position of the flow sensor. The alignment pin must fit in the recess
on the gauge mounting block.
•
Tighten the red coloured retainer nut by hand. Tightening by hand should be sufficient. However, if
the seal is not leak tight carefully tighten the nut with an appropriate tool a bit further
•
The mechanical installation of the flow sensor is therewith completed.