E+E Elektronik EE75 User Manual Download Page 6

5

Hardware

3.3  Mounting Type T3 (Remote Probe)

3.3.1  Installing the enclosure

 

Refer to chapter 3.1

3.3.2  Installing the probe

 

Refer to chapter 3.2.2 - Mounting with a Flange

3.4  Mounting Type T26 (Remote Probe, Pressure- 

 

tight up to 10 bar 

/ 145 psi

)

3.4.1  Installing the Enclosure

 

Refer to chapter 3.1

3.4.2  Installing the Probe

General safety instructions for installation using pressure-

tight movable screw connections

 

As the sensing probe can be exposed to very high pressures in the 

measuring environment, there is a risk of sudden, unintentional 

expulsion of the probe during or after improper installation. Special 

care should therefore be taken when working on or in the vicinity of the 

sensing probe. Do not, under any circumstances, bend directly over the 

probe. 
When installing the sensor probe, be careful not to damage the surface 

of the sensing probe. 
This could damage the seals (resulting in leakage and pressure loss) 

or cause problems when removing the probe (may get stuck).
Before installing the sensing probe, make sure it is free from any 

impurities such as grease or dirt.

Installing the probe

There should be a shut-off valve on both sides of the probe insert when 

installing the probe. This makes it easy to remove the sensor for 

maintenance and calibration.

1. Install the probe with the shut-off valves closed.

2. Insert the sensor probe into the process.

3. To ensure the probe is installed securely, the lock nut must be  

  tightened to a specified torque of 10 Nm.

  If no torque spanner is available, tighten the lock nut by hand as far  

  as possible and then turn it a further ~ 90° with an appropriate  

  open-ended spanner.

An inadequate torque results in a low tension force (fixing force) 

on the clamping sleeve. This brings with it a risk of injury due to 

sudden expulsion of the sensing probe. An excessive torque can 

lead to permanent deformation of the clamping sleeve and 

sensing probe, making removal and re-installation more difficult 

or even impossible.

Selected probe length

12

(0.5")

8

(0.3")

Selected probe length

12

(0.5")

8

(0.3")

shut-off
valve

shut-off 
valve

clamping 
ring

1/2" ISO  

or NPT

lock nut

sealing ring

bore 

> 9.2mm (0.4")

Summary of Contents for EE75

Page 1: ...BA_EE75 v14 Modification rights reserved 192310 User Manual EE75 Air Gas Velocity Sensor Hardware and Software...

Page 2: ...and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that inte...

Page 3: ...Calibration 10 6 1 2 General Information on 2 point v T Calibration 11 6 2 Velocity Calibration 11 6 2 1 Calibration Procedure Using Configuration Software 11 6 2 2 Calibration Procedure Using Button...

Page 4: ...ctions Excessive mechanical stress and improper use must be avoided As the sensor element is sensitive to electrostatic discharge ESD appropriate protective measures should be taken when touching it I...

Page 5: ...rt of the enclosure is fitted using 4 screws not included Max screw diameter 4 5 mm 0 18 for example 4 2 x 38 mm 0 17 x 1 5 DIN 7983H screws 4 Connecting the sensor see chapter 4 Electrical Connection...

Page 6: ...e may get stuck Before installing the sensing probe make sure it is free from any impurities such as grease or dirt Installing the probe There should be a shut off valve on both sides of the probe ins...

Page 7: ...centre of the velocity measuring slot middle of the sensor The scaling on the probe has an offset of 30mm 1 2 This allows for the insertion depth X distance between middle of sensor and mounting plat...

Page 8: ...ssignment USB Interface Cable Colour of wire Signal Terminal no Green D 11 White D 12 Red Vbus 13 Black USB GND 14 Shielding Description Connection assignment VCC 5 GND 4 GND 3 OUT2 2 OUT1 1 Descripti...

Page 9: ...ing jumpers Jumper position Also see chapter 5 1 Circuit Board Response time No jumper 1 5 s Jumper at pos 1 2 s factory setting Jumper at pos 2 4 s factory setting Jumper at pos 3 10 s factory settin...

Page 10: ...ted 6 Status Line The status line indicates the sensor s current operating status no indication Normal operating mode measurement in progress MIN or MAX See point 4 MIN MAX function CAL LOW V or T cal...

Page 11: ...mited working ranges 2 point v T calibration With 2 point calibration precise measuring results can be obtained over the entire v T measuring range The more complicated 2 point calibration procedure i...

Page 12: ...described in chapter 6 1 1 If calibration point 50 of measuring range select V CAL HIGH and confirm the selection If calibration point 50 of measuring range select V CAL LOW and confirm the selection...

Page 13: ...adjusted measuring value 10 In the case of 1 point T calibration the procedure is now complete 11 For 2 point T calibration repeat steps 6 2 9 with the high calibration point 6 4 Resetting to Factory...

Page 14: ...120 C 40 248 F Accuracy typ 2 0 5 C 0 9 F in air at 25 C 77 F Response time t90 typ 10 s Temperature dependency electronics typ 0 005 of mv K deviating from 25 C 77 F mv measured value v or T Temperat...

Page 15: ...The permanent connection between EE75 and PC via the USB interface in normal operation may cause a malfunction and does not meet the CE criteria 2 Installation NOTE Administrator authorisations may be...

Page 16: ...ntroller on your PC s Control Panel See Start Settings Control Panel System Hardware Device Manager 6 The configuration software can now be opened by double clicking the EE75 icon on your desktop 7 Sp...

Page 17: ...Start Settings Control Panel System Hardware Device Manager Closes the configuration software 3 2 Help Provides general information on the configuration software 4 EE75 Configurator 4 1 Start The Sta...

Page 18: ...nput field Single line display Two line display factory setting Backlight on check box Activated ON Not activated OFF 4 4 Response Time As described under Hardware chapter 5 2 Jumpers for Setting the...

Page 19: ...ld If the duct has a different shape or the cross sectional area is already known the value can of course also be entered directly in the Cross sectional area field in m or ft 4 6 Probe Cable The leng...

Page 20: ...he media pressure is significantly different the actual values can be defined in the input field on the Mediacorr tab to obtain the best possible measuring results 4 9 Calibration Velocity and tempera...

Page 21: ...HIGH should be around 1 8 m s 1 4 2 4 9 3 V Calibration Procedure Using the Configuration Software 1 Position the sensor head in the middle of the reference system wind tunnel 2 Set the required calib...

Page 22: ...particularly useful for checking sensors which do not have the optional display module 4 11 Information The Information tab contains general information on the sensor It also saves the date of the la...

Page 23: ...22 Configurations Software...

Page 24: ...pluse fr E E Elektronik Germany Obere Zeil 2 61440 Oberursel Tel 49 6171 69411 0 E mail info epluse de E E Elektronik India 801 Sakhi Vihar Road 400072 Mumbai Tel 91 990 440 5400 E mail info in epluse...

Reviews: