User Manual EE371 Compact Dew Point Senor

1  General

This user manual serves for ensuring proper handling and optimal functioning of the device. The user 

manual shall be read before commissioning the equipment and it shall be provided to all staff involved 

in transport, installation, operation, maintenance and repair. The user manual may not be used for the 

purposes of competition without the written consent of E+E Elektronik® and may not be forwarded to 

third parties. Copies may be made for internal purposes. All information, technical data and diagrams 

included in these instructions are based on the information available at the time of writing.

Please find this document and further product information on our website at  

www.epluse.com/ee371.

Disclaimer of Liability

The manufacturer or his authorized agent can be only be held liable in case of willful or gross 

negligence. In any case, the scope of liability is limited to the corresponding amount of the order issued 

to the manufacturer. The manufacturer assumes no liability for damages incurred due to failure to 

comply with the applicable regulations, operating instructions or the specified operating conditions. 

Consequential damages are excluded from any liability.

1.1  Explanation of Symbols

This symbol indicates safety information.

It is essential that all safety information is strictly observed. Failure to comply with this information can 

lead to personal injuries or damage to property. E+E Elektronik® assumes no liability if this happens.

This symbol indicates instructions.

The instructions shall be observed in order to achieve optimal performance of the device.

1.2  Safety Instructions

1.2.1  General Safety Instructions

 

ƒ

Avoid any unnecessary mechanical stress and inappropriate use.

 

ƒ

When replacing the filter cap make sure not to touch the sensing elements.

 

ƒ

The device must be operated with the filter cap on at all times.

 

ƒ

For sensor cleaning and filter cap replacement please see the “Cleaning Instructions” at  

www.epluse.com.

 

ƒ

Installation, electrical connection, maintenance and commissioning shall be performed by qualified 

personnel only.

 

ƒ

Use the EE371 only as intended and observe all technical specifications.

 

ƒ

Do not use EE371 in explosive atmosphere or for measurement of aggressive gases.

 

ƒ

Do not apply the nominal voltage to the data lines.

1.2.2  Intended Use

The EE371 is intended for the dew point temperature (Td) measurement of compressed air and other 

non-corrosive and non-flammable gases at pressures. The sensor can be installed in a pressurized 

system up to 20 bar (290 psi) and 100 bar (1 450 psi), respectively. 

The use of the EE371 in any other way than described in this manual bears a safety risk for people and 

the entire measurement installation and is therefore not allowed. 

The manufacturer cannot be hold responsible for damages as a result of incorrect handling, installation, 

and maintenance of the equipment.

In order to avoid damage to the instrument or health hazards, the measuring equipment must never be 

manipulated with tools that are not specifically described in this manual.

The sensor may only be utilized in accordance with the conditions defined in the technical data. 

Otherwise, inaccuracies of the measurement will occur and equipment failures cannot be ruled out.

The steps recommended by the manufacturer for installation, inspections and maintenance work must 

be observed and carried out for the safety of the user and for the functionality of the equipment.

Summary of Contents for EE371

Page 1: ...BA_EE371 v2 1 Modification rights reserved User Manual EE371 Compact Dew Point Sensor...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...7 4 1 Installation Location 7 4 2 Installing Directly into the Process 8 4 3 Mounting with a Sampling Cell 8 4 3 1 Basic Sampling Cell 9 4 3 2 Sampling Cell with Quick Connector 9 5 Electrical Connec...

Page 4: ...ice 1 2 Safety Instructions 1 2 1 General Safety Instructions Avoid any unnecessary mechanical stress and inappropriate use When replacing the filter cap make sure not to touch the sensing elements Th...

Page 5: ...not put damaged products into operation and protect them from accidental commissioning Mark the damaged product as defective A faulty device may only be investigated and possibly repaired by qualifie...

Page 6: ...power on This advanced auto calibration mechanism takes a little bit longer than the regular auto calibration and is performed up to 5 times during the first hour of operation During auto calibration...

Page 7: ...should be paid to the fact that changes in the pressure of a gas also changes the dew point If there is a pressure difference between the location of installation and the process the measurement can b...

Page 8: ...too high Sensor shall be protected against contamination Removing the sensor must not interrupt the process To obtain a representative sample of the process gas and to avoid measuring errors please no...

Page 9: ...sic Sampling Cell Pressure range 0 64 bar 0 928 psi Order code HA050103 ISO HA050105 NPT 1 2 3 ISO NPT 1 G 1 2 1 2 2 G 1 4 1 4 3 G 1 4 1 4 Fig 6 Basic sampling cell 4 3 2 Sampling Cell with Quick Conn...

Page 10: ...C V mA V mA 10 30V DC 1 7 6 5 2 3 4 1 V 2 GND 3 NC11 4 NC12 5 NC21 6 NC22 7 not connected 10 30V DC Analogue output The switching thresholds are factory set 6 Operating Components 6 1 Electronics Boar...

Page 11: ...EE PCS Product Configuration Software and the optional RS232 interface cable order code HA010304 or with the EE PCA Product Configuration Adapter and the according connection cable order code HA011063...

Page 12: ...ch as rubbing or brushing leads certainly to their irreversible damage Please refer to the Cleaning Instructions available online at the E E Download Center 8 2 Filter Replacement A contaminated filte...

Page 13: ...e and scaleable 0 10 V 0 mA IL 1 mA analogue outputs for Td Tf Wv 4 20 mA 3 wire RL 500 Ohm RL load resistance General Power supply class III 2 15 30 V DC Current consumption typ voltage output 40 mA...

Page 14: ...92310 S vres Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 19 20...

Reviews: