User Manual EE371 Compact Dew Point Senor

9  Accessories

Description

Order code

EE-PCS Product Configuration Adapter

EE-PCA

Connection cable for connecting the EE371 with EE-PCA

HA011063

RS232 interface cable for connecting the EE371 with EE-PCS

HA010304

Sampling cell G 1/2" 

with quick connector

HA050102

Sampling cell NPT 

with bleed screw

HA050107

Sampling cell G 1/2" 

for atmospheric dew point

HA050106

Basic sampling cell G 1/2"

HA050103

Basic sampling cell NPT

HA050105

10  Technical Data

Measurands

Dew point (Td)

Measuring range  

  -60...60 °C Td 

(-76...140 °F Td)

Accuracy

1) 

 

80

60

40

20

0

-20

-40

-60

0

10

20

30

40

50

60

-10

-20

-30

-40

80

60

40

20

0

-20

-40

-60

-80

0

10

20

30

40

-10

-20

-30

-40

60

50

Accuracy ≤ ±2 °C Td (≤ ± 3.6 °F)

Process temperature [°C]

D

ew poin

t t

emper

atur

e [°C

]

 

 

 

 

 

 

 

 

Response time t

90

  

80 s  -20 °C Td 

   -40 °C Td 

(-4 °F Td 

→  

-40 °F Td) 

  

10 s  -40 °C Td 

   -20 °C Td 

(-40 °F Td 

   -4 °F Td)

Volume concentration

Measuring range  

  20...200 000 ppm

Accuracy at 20 °C 

(68 °F)

 and 

  ±(5 ppm + 9 % from measured value) 

1 013  mbar 

(14.7 psi)

Outputs

Two freely selectable and scaleable  

  0 - 10 V 

         0 mA < I

L

 < 1 mA 

analogue outputs for Td, Tf, Wv 

  4 - 20 mA, 3-wire           R

L

 < 500 Ohm

    

R

L

 = load resistance

General

Power supply class III 

2)

 

  15 - 30 V DC

Current consumption, typ.  voltage output  40 mA / during auto-calibration: 100 mA  

at 24 V DC 

current output  80 mA / during auto-calibration: 140 mA

Pressure range 

  0...20 bar 

(0...290 psi)

 / 0...100 bar

 (0...1 450 psi)

Enclosure material  

  Al Si 9 Cu 3

Protection rating 

  IP65

Electrical connection 7-pole industrial plug  DIN VDE 0627 / IEC 61984 

 

wire cross-section  0.25 - 1 mm

2

 

 

cable outlet  PG 11

Sensor protection 

  Stainless steel sintered filter

Working temperature range medium (air):  -40...70 °C 

(-40...158 °F)

 

 

electronics:  -40...60 °C 

(-40...140 °F) 

 

display:  -20...50 °C 

(-4...122 °F)

Storage temperature range 

  -40...60 °C 

(-40...140 °F)

Electromagnetic compatibility 

  EN 61326-1  EN 61326-2-3   

 

  Industrial Environment  

 

  FCC Part15 ClassB  ICES-003 ClassB

1) Traceable to international standards, administrated by NIST, PTB, BEV... 

  The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was  

  calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
2) Class 2 in USA & Canada

Summary of Contents for EE371

Page 1: ...BA_EE371 v2 1 Modification rights reserved User Manual EE371 Compact Dew Point Sensor...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...7 4 1 Installation Location 7 4 2 Installing Directly into the Process 8 4 3 Mounting with a Sampling Cell 8 4 3 1 Basic Sampling Cell 9 4 3 2 Sampling Cell with Quick Connector 9 5 Electrical Connec...

Page 4: ...ice 1 2 Safety Instructions 1 2 1 General Safety Instructions Avoid any unnecessary mechanical stress and inappropriate use When replacing the filter cap make sure not to touch the sensing elements Th...

Page 5: ...not put damaged products into operation and protect them from accidental commissioning Mark the damaged product as defective A faulty device may only be investigated and possibly repaired by qualifie...

Page 6: ...power on This advanced auto calibration mechanism takes a little bit longer than the regular auto calibration and is performed up to 5 times during the first hour of operation During auto calibration...

Page 7: ...should be paid to the fact that changes in the pressure of a gas also changes the dew point If there is a pressure difference between the location of installation and the process the measurement can b...

Page 8: ...too high Sensor shall be protected against contamination Removing the sensor must not interrupt the process To obtain a representative sample of the process gas and to avoid measuring errors please no...

Page 9: ...sic Sampling Cell Pressure range 0 64 bar 0 928 psi Order code HA050103 ISO HA050105 NPT 1 2 3 ISO NPT 1 G 1 2 1 2 2 G 1 4 1 4 3 G 1 4 1 4 Fig 6 Basic sampling cell 4 3 2 Sampling Cell with Quick Conn...

Page 10: ...C V mA V mA 10 30V DC 1 7 6 5 2 3 4 1 V 2 GND 3 NC11 4 NC12 5 NC21 6 NC22 7 not connected 10 30V DC Analogue output The switching thresholds are factory set 6 Operating Components 6 1 Electronics Boar...

Page 11: ...EE PCS Product Configuration Software and the optional RS232 interface cable order code HA010304 or with the EE PCA Product Configuration Adapter and the according connection cable order code HA011063...

Page 12: ...ch as rubbing or brushing leads certainly to their irreversible damage Please refer to the Cleaning Instructions available online at the E E Download Center 8 2 Filter Replacement A contaminated filte...

Page 13: ...e and scaleable 0 10 V 0 mA IL 1 mA analogue outputs for Td Tf Wv 4 20 mA 3 wire RL 500 Ohm RL load resistance General Power supply class III 2 15 30 V DC Current consumption typ voltage output 40 mA...