![E+E Elektronik EE310 Operation Manual Download Page 32](http://html1.mh-extra.com/html/e-e-elektronik/ee310/ee310_operation-manual_3078795032.webp)
32
Operating instructions for EE310 Humidity / Temperature Transmitter
5.4 Configuration Menu
Detailed information to the configuration menu see chapter
10 Appendix
.
Menu
Data-logging
Display settings
Analog output
Customer
adjustment
Alarm output*
Modbus
settings*
configuration of data logger/graph
- sampling rate | graphs
setting of display layout
- measurands | brightness | display alarm
output configuration
- mode | measurands | scaling | error indication
relay configuration
- mode | set points | state
adjustment -
1 and 2 point humidity/temperature adjustment |
reset to factory adjustment | calibration status
configuration of Modbus RTU data transmission
Device settings
Status
settings
- language | date, time | parameters | password protection
status and device information
IP settings*
configuration of Ethernet Module
* Menu only available with connected corresponding modules.
5.5 Status information
i
The status information shows all actual EE310 settings.
Fig. 33
Status information
5.6 Buzzer ON / OFF
Buzzer ON
Buzzer OFF
5.7 Error indication
When an error occurs, the error indication shows the error code.