P/N: 975389
|
Rev: current
|
ECO: xxxxx
www.dynisco.com
From lab to production,
providing a window into the process
-32-
•
Feed power cable thru the cable grommet in the bottom right side of the RCU
cabinet and clamp in place for strain relief
•
Connect the ground wire to terminal 2 of terminal block 5 (TB5).
•
Connect the line wire to terminal 2 of the ON/OFF switch (S1).
•
Connect the neutral wire to terminal 4 of the ON/OFF switch (S1).
•
Verify all connections are secure and tight.
•
Verify all breakers are turned to the on position.
•
Close and lock cabinet prior to turning on the system.