Connectors, Cable glands, Cable
10. Troubleshooting
10.1 Troubleshooting
Fault
No signal
Strong zero shift when screwing
in
Mounting hole incorrectly
produced (alignment error)
Mounting torque too high
Check hole with test bolt (chapter 6.2),
rework with tool if necessary
Adjust to max. 50 Nm mounting torque
Check mounting hole; remove solidified
plastic
Send pressure transmitter to
DYNISCO
for repair
Plug forming in front of
diaphragm
Diaphragm damaged
No signal change despite
pressure rise
Possible cause
Cable breakage or poor contact
No supply voltage
Resolution
Check cable and contact, or
replace
Check supply voltage